Add Condition:

Product Description
Specifically designed for the long-term home use, the Respironics Lifecare PLV-100 is a reliable, easy to use portable volume ventilator. The PLV-100 is a versatile device that's effective on a wide range of patients including pediatric patients and patients with minute ventilation demands.
FEATURES:
Wide Range Of Applications:
- Use in home, sub-acute care andchronic care settings for transport
- Small size and light weight ideal for mobile patients in wheelchairs, cars, and RVs
- Capable of ventilating pediatric to adult patients
- O2 bleed-in up to 40%
- Provides volume or pressure ventilation
Digital Displays:
- Tidal Volume
- I:E Ratio
- Respiratory Rate
- Inspiratory Flow Rate
Alarm Interface Capabilities:
- Remote Alarm
- Pressure Alarm
- Hospital/Nurse Call System
Three Modes Of Operation:
- Control l Assist-Control
- Spontaneous Intermittent Ventilation (SIMV)
Multiple Power Sources:
- 120 VAC/240 VAC
- 12 VDC external capabilities (up to 20 hours)
- Internal battery (1 hour fully charged)
SPECIFICATIONS
Front Panel Controls
- Mode Selection: Control, Assist/Control, SIMV
- Tidal Volume: 0.05 to 3.00 liters ± 10%
- Breath Rate: 2 to 35 BPM ± 0.5; 36 to 40 BPM ± 2
- Inspiratory Flow Rate: 10 to 120 L/Min
- Battery Test Switch: Test Internal/External Battery
- Main Power Switch: On/Off, Recharge
- DC Circuit Breaker: Push/Pull Actuation
- AC Circuit Breaker: Push/Pull Actuation
- Airway Pressure Limit: 5 to 100 cm H2O ± 5 cm H2O
- Patient Airway Pressure: -10 to 100 cm H2O ± 5
- Sensitivity: +3 cm H2O ± 1 to at least -6 cm H2O
- Low Pressure Alarm: 2 to 40 cm H2O ± 2; 41 to 50 cm H2O ± 5
Alarms
The audible alarm ranges between 70 and 85decibles (at 1 meter) and has its own power sourceseparate from the internal battery. The power source and batteries are constantly recharged with a 2 amp trickle charge when the unit is plugged into AC power, whether it is operating or not.
- Low Pressure: Visual/Continuous Audible
- High Pressure: Intermittent Audible
- Apnea: Visual/Continuous Audible
- Inverse I:E Ratio: Visual Warning
- IncreasedInspiratory Flow: Visual Warning
- Low Internal Battery: Visual/Continuous Audible
- Low External Battery: Visual/Continuous Audible
- Reverse Battery Connection: Continuous Audible
- Power Failure: Continuous Audible
- Power Source: Audible
- Unit Malfunction: Intermittent Audible
TOP^Recently Viewed Products