Description

Ultra-lightweight and portable with up to 6 hours of battery life.

Measures Non-Invasive Blood Pressure, SpO2, Stat-Temperature.

SureTemp™ Thermometer provides an accurate reading within seconds

  • Suitable for adult, pediatric and neonatal patients
  • Nellcor SpO2 technology, and motion tolerant SPO2 and NIPB
  • Adjustable audible and visual alarms
  • Removable and rechargeable battery provides up to 6-hour working time
  • Integrated handle, full mounting solutions and integrated printer

Specifications

Physical Weight & Dimensions

  • Height: 240 mm
  • Width: 170 mm
  • Depth: 170 mm
  • Weight: < 3.5g (including a battery)

Safety

Meets the requirements of IEC60601 series, CE marking according to MDD93/42/EEC

Operation Environment

  • AC Power Req: 100-240VAC (± 10%), SO/60 Hz (± 3 Hz), 70VA
  • Temperature: 0-40°C (SureTemp™: 10-40°C)
  • Humidity: 15%-95%, non-condensing

Patient Range

Adult, pediatric and neonatal

NIBP

  • Method: Automatic oscillometric
  • Operation modes: Manual / Automatic / Continual
  • Auto measure time: Adjustable
  • Measurement unit: mmHg / kPa selectable
  • Measurement types: Systolic, Diastolic, Mean, Pulse Rate
  • Measurement range: Adult: 10-270mmHg; Pediatric: 10-200mmHg; Neonate: 10-135mmHg

Over-pressure

  • Protection: YES
  • Resolution: 1mmHg
  • Alarm: Systolic, Diastolic, Mean, Pulse Rate
  • Pulse Rate Range: 40-240 bpm
  • Pulse Rate Resolution: 1bpm
  • Pulse Rate Accuracy: ± 2bpm

Pulse Rate

Range and accuracy: 20-250 bpm ± 3 bpm, 251-300 bpm unspecified

Typical Measurement

Time: < 12s

SpO2

  • Measurement range: 0-100%
  • Resolution: 1%
  • Accuracy:
    • Adult/Pediatric: ±2% (70-100%)
    • Neonatal: ±3% (70-100%)
    • 0-69% unspecified
  • Alarm range: 0-100%
  • Pulse rate:
    • Range: 20-254bpm
    • Resolution: 1bpm
    • Accuracy: ± 3bpm
    • Alarm range: 20-254bpm
    • Storage Humidity: 10%RH to 95%RH

Nellcor OxiMax SpO2

  • Measurement range: 0-100%
  • Compatible sensors:
    • Nellcor sensor type: OxiMax series MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I, MAX-FAST, MAX-R,OxiCliq A, OxiCliq N, OxiCliq P, OxiCliq I, D-YS, D-YSE, D-YSPD, DS-100A, OXI-AIN and OXI-P/I
    • SpO2 accuracy:
      • 70%-100% ±2%; 0-69% unspecified (sensors: MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I and MAX-FAST)
      • 70%-100% ± 2.S%; 0-69% unspecified (sensors: OxiCliq A, OxiCliq N, OxiCliq P and OxiCliq I)
      • 70%-100% ± 3%; 0-69% unspecified (sensors: D-YS, DS-100A, OXI-A/N and OXI-P/I)
      • 0%-100% ±3.5%; 0-69% unspecified

SureTemp™ Thermometer

  • Measurement range: 25 - 44°C (77-111.2°F)
  • Resolution: In MONITOR mode: 0.1°C (0.2°F)
  • Accuracy:
    • In MONITOR mode: 25-32°C (77-89.6 °F) : ±0 .2°C (±0.3 °F) including 32°C (89.6 °F)
    • In MONITOR mode: 32~44°C (89.6-111.2 °F) : ±O.l °C (±0.2 °F) excluding 32°C (89.6 °F)