DRE Medical Equipment

(800) 462-8195

1.502.244.4444

This item has been identified as hard to find, discontinued, or unavailable at this time. Although we will be happy to work with you in locating this item, please consider the alternatives listed below.

Refurbished - CASMED 740 Vital Signs Monitor

Condition:
  • Refurbished
Market Availability:
Scarce
download sales sheetSales Sheet
More Resources
Manuals, Docs, Links
Casmed 740 Vital Signs Monitor

Product Description

The CASMED® 740 monitor offers leading-edge technology in a rugged, portable, lightweight unit that weighs only 1.4 kg (3 lbs).

Easy to use, reliable, accurate, with multi-parameter capabilities

  • Fast, accurate blood pressure readings for neonates through adults . . . reduces the problems of motion in blood pressure measurement
  • Monitor oximetry and temperature either continuously or in a “spot-check” mode
  • Ideal for medical/surgical patients on pain medications including PCA pumps, endoscopy procedures, and for the EMS market

 

Featured technologies

  • CASMED’s proprietary MAXNIBP® technology

Oximetry choices include:

  • Masimo SET®
  • Nellcor® OxiMax®
  • Nonin®

Four models available:

  • CAS 740-1: MAXNIBP
  • CAS 740-2: MAXNIBP and SpO2
  • CAS 740-2T: MAXNIBP and Temperature
  • CAS 740-3: MAXNIBP, SpO2 and Temperature

 

FEATURES

  • Created with emergency medical responders in mind
  • Wireless (infrared) printer option
  • User removable NiMH battery pack, internal power supply and DC power option
  • Flexible mounting solutions include cart, IV pole, bedrail, or wall; it can be fixed to a swivel wall mount designed specifically for emergency vechicles or carried in a protective, durable case

 

TECHNICAL SPECIFICATIONS

NIBP

  • Technique: Oscillometric (MAXNIBP® Technology)
  • Patient Range: Adult – Neonate
  • Measurement Range
    - Systolic: 30-255 mmHg
    - Diastolic: 15-220 mmHg
    - MAP: 20-235 mmHg
  • Pulse Rate Range: 30-240 BPM
  • Accuracy:
    - Blood Pressure: ± 5 mmHg standard deviation no greater than 8 mmHg
    - Pulse Rate: ± 2% or ±2 BPM, whichever is greater

 

Oximetry (Options)

  • Masimo SET®
    - SpO2 Range: 0-100%
    - SpO2 Accuracy:
        70-100%, ± 2 digits (1 S.D.) Adult
        70-100%, ± 3 digits (1 S.D.) Neonate
    - Pulse Rate Range: 25-240 BPM
    - Pulse Rate Accuracy: ± 3 BPM
  • Nonin®
    - SpO2 Range: 0-100%
    - SpO2 Accuracy:
        70-100%, ± 2 digits (1 S.D.) Adult
        70-100%, ± 3 digits (1 S.D.) Neonate
    - Pulse Rate Range: 18-240 BPM
    - Pulse Rate Accuracy: ± 3% or ±1 digit, whichever is greater
  • Nellcor® OxiMax®
    - SpO2 Range: 1-100%
    - SpO2 Accuracy:
        70-100%, ± 2 digits (1 S.D.) Adult
        70-100%, ± 3 digits (1 S.D.) Neonate
    Pulse Rate Range: 20-240 BPM
    Pulse Rate Accuracy: ± 3 digits

Temperature (Optional)

  • Welch Allyn SureTemp® Plus
    - Temperature Range: 29 °C to 42 °C (84.2 °F to 107.6 °F)
    - Accuracy: ± 0.1 °C (± 0.2 °F)
    - Meets or exceeds ASTM Standards

Control Panel

  • Display: LED display of measurement results, instructions, troubleshooting messages, and signal strength bar
  • Parameters Displayed: Systolic Pressure, Diastolic Pressure, Mean Arterial Pressure (MAP), Pulse Rate, %SpO2, Temperature (°C or °F)

Patient Alarms

  • Adjustable high and low alarms for Pulse Rate, SpO2, Systolic Pressure, Diastolic Pressure and Mean Arterial Pressure

Operating Modes

  • History: Review of previous measurements
  • NIBP: Manual, STAT or Automatic (at preset intervals)
  • %SpO2: Continuous Monitoring
  • Temperature: Predictive or Continuous

Power

  • Source: External line or internal battery
  • AC Power: 100-240 VAC, 50-60 Hz, 0.5A; T1.25A L250V fuse rating
  • DC Power (EMS Option): 12-30 VDC
  • Battery: NiMH battery pack (removable)
  • Charge Time: 4 hours
  • Operating Time: Minimum 4 hours of continuous operation
  • Leakage Current: 100 _A (maximum)

Safety Limits

  • 145 mmHg (Neonate)
  • If measurement time exceeds 120 seconds (Adults), 90 seconds (Neonate)
  • If safety timer detects microprocessor failure

Operating Environment

  • Operating Temperature: 0° C to 50° C (32° F to 122° F)
  • Storage Temperature: -20° C to 65° C (-4° F to 149° F)
  • Humidity: 15 - 95%, non-condensing
  • Altitude: Up to 3048 meters (10,000 feet) if stored or used outside temperature and humidity ranges.

Physical Dimensions & Weight (Base Unit)

  • H x W x D: 17.0 cm x 21.5 cm x 7.5 cm
  • (6.75 in x 8.5 in x 3.0 in)
  • Weight: 1.4 kgs (3 lbs)

Standards

  • ANSI/AAMI SP-10
  • EN 60601-1, EN 60601-1-2, EN 865, EN 60601-2-30, EN 60601-2-49
  • ETL Listed – UL 60601, CAN/CSA C22.2 No. 601.1
  • CE Marking according to Directive 93/42/EEC

Optional Accessories

  • Infrared printer
  • RS-232 interface
  • Swiveled hard mount (for ambulance applications)
  • Roll stand and basket
  • Protective carrying case

Suggested Alternatives

TOP^

Helpful Resources