DRE Medical Equipment

(800) 462-8195

1.502.244.4444

Tidalwave 715Sp Handheld Monitor with SP O2

Model Number:
600715
Condition:
  • Retired
Market Availability:
Limited
More Resources
Manuals, Docs, Links
Tidalwave 715Sp Handheld Monitor with SP O2
Tidalwave 715Sp Handheld Monitor with SP O2

Product Description

DRE offers the Respironics Novametrix Tidal Wave model 715Sp portable handheld pulse oximeter and CO2 monitor with SpO2.  The Tidal Waves are the only handheld, mainstream capnographs available today, and provide immediate feedback on the patient's airway condition and ventilatory status. Based on the same technology as Respironics Novametrix bedside capnographs.

THREE OPTIONAL CONFIGURATIONS:

  • Tidal Wave Model 610
    Handheld Mainstream Capnograph

     
  • Tidal Wave Model 615
    Handheld Capnograph with Mainstream and Sidestream Sampling

     
  • Tidal Wave Sp Model 710
    Handheld Mainstream Capnograph/Pulse Oximeter Tidal Wave Sp

     
  • Tidal Wave Sp Model 715 (described on this page)
    Handheld Mainstream Capnograph/Pulse Oximeter with Sidestream Sampling Capabilities

FEATURES:

  • Built in Pulse Oximeter
  • Hand-held Mainstream and Sidestream Capnograph
  • 30-minute on-screen graphic trend available
  • Visual and audible “no respiration” alert
  • User replaceable disposable and rechargeable battery packs eliminate the need to wait while batteries are charging (standard AA lithium: 5.5 hours of life)
  • No calibration required for CO2

CAPNOGRAPH

Principle of Operation
Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.

Initialization Time
Capnogram within 15 seconds, full specifications within 60 seconds.

Calibration
No calibration required.

CAPNOSTAT® Mainstream CO2 Sensor
Size: 3.30 × 4.24 × 2.16 cm (1.30 × 1.67 × 0.85 in.).
Weight: 18 grams (0.64 oz.), cable excluded
Shock Resistance: Withstands repeated six foot drops to a hard surface.

CO2 (Carbon Dioxide)
Range: 0 - 150 mmHg
Accuracy:
   ±2 mmHg (for 0 - 40 mmHg),
   ±5% of reading (for 41 - 70 mmHg),
   ±8% of reading (for 71 - 150 mmHg)
Resolution: 1 mmHg
Response Time: 60 ms
Units: mmHg, kPa, or %
Compensations: N2O & O2 (selectable), Pb (automatic)
Capnogram: Selectable sweep speed

Respiratory Rate
Accuracy: ±1 br/min
Range: 0-150 br/min
Resolution: 1 br/min

Sampling Flow Rate (Model 715 only)
180 ml/min (for non-intubated patients)
Response time is less than 200 ms at 180 ml/min flow

PULSE OXIMETER

Principle of Operation
Red/Infrared absorption.

SpO2 (Oxygen Saturation)
Range: 0-100%
Accuracy: ±2% SpO2 (for 80-100% SpO2)
(1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
Resolution: 1%
Averaging: Menu selectable 2 or 8 seconds

Pulse Rate
Range: 30-250 bpm
Accuracy: ±1% of full scale
Resolution: 1 bpm
Averaging: 8 seconds

Sensors
Reusable Y-Sensor (can be sterilized and used with all patient populations);
reusable adult finger sensor

GENERAL

Alerts
Limits:Adjustable limits for ETCO2, SpO2, Respiratory Rate and Pulse Rate. No Respiration Detected adjustable timeout.
Audio: Adjustable volume, two minute silence or off (LED indicators)
Visual: On-screen and LED

Display
Type: Dot matrix, backlit (LCD)
Size: 2.2 × 1.1 inches

Trending Capabilities
Memory: 24 hr., battery backed
Format: 30 minute on-screen or printed in trend or tabular mode.
  Download data to personal computer for use with NovaCARD software

Communication Output
RS232 to Seiko DPU-414 printer, NovaCARD software or computer

Physical
Size (height, width, depth): 20 × 9.5 × 4.8 cm (8 × 3.75 × 1.875 in.).
Weight: 793 grams (28 oz.).

Electrical
External Supply (optional)
Input: 120 VAC 60 Hz, or 220 VAC 50 Hz
Battery: Nickel-Metal-Hydride (NiMH): 4.5 hour life
Battery: Standard AA lithium: 5.5 hour life

Environmental
Operating temperature: 0 to +40 °C (32-104 °F)
Operating humidity: 0 to 90% RH non-condensing.