DRE Medical Equipment
englishespañol
phone: 1-800-462-8195DRE Facebook PageDRE LinkedIn PageDRE Twitter Profile

Refurbished - Corometrics 118 Fetal Monitor

Condition:
  • Refurbished
Market Availability:
Limited
Refurbished - Corometrics 118 Fetal Monitor

Product Description

The Corometrics 118 Fetal Monitor is a twin capable unit that includes one toco and two ultrasound transducers, fetal EKG, event marker and a chart recorder.  Data storage and transmission Integrated Maternal Pulse Oximetry NIBP.  Extends upon the advanced technology of the Model 116 with integrated maternal pulse oximetry and/or noninvasive blood pressure monitoring.

Features:

  • Easy operation
  • Intuitive controls and menus
  • Simplified operation of all functions
  • Controls are tailored for labor and delivery to make operation user friendly
  • Visual/Audible Maternal Alarms
  • The unit provides visual and audible alarms when maternal SpO2, NIBP or pulse rate values exceed user-defined high/low limits
  • Audible alarms can be disabled
  • Individual audio/visual alarms can be temporarily cancelled

Specifications:

Direct ECG

  • Technique:  peak detecting beat-to-beat cardiotachometer. 
  • Heart rate counting range: 30-240 BPM. 
  • Heart rate resolution: ±1 BPM. 
  • Artifact elimination:  switch selectable ±25 BPM artifact rejection. 
  • Input signal range differential:  15 µV to 2 mV peak-to-peak (fetal), 100 µV to 2 mV peak-to-peak (maternal). 
  • Common mode: 20 VDC (max.). 
  • Preamplifier bandwidth: 1.0 Hz to 100 Hz. 
  • Common mode rejection: >120 dB balance. 
    With patient cable: >110 dB 5k RA or LA unbalance. 
  • Input equivalent noise: < 10 µV peak-to-peak. 
  • Input impendence:  differential mode > 10 MΩ, common mode > 20 MΩ.
  • Main frequency rejection: >40 dB. 
  • Leakage current:  < 10 µA at 120 VAC. 
  • Isolation mains to patient connection:  >2kV. 
  • Leg plate tester: simulated r-wave at 120 ±1 BPM.
Ultrasound
  • Technique:  pulsed doppler with autocorrelation processing.
  • Pulse repetition frequency:  4kHz in single ultrasound mode, 2 kHz in dual ultrasound mode.
  • Pulse duration:  92µs.
  • Center frequency:  1.151 MHz
  • Intensity:  <10 mW/cm2
  • Heart rate counting range:  50-210 BPM.
  • Leakage current: <10 µA at 120 VAC.
Pressure
  • Range: 0-100 mmHg
  • Tocotransducer: 0-100 relative units (0-120 grams)
  • Excitation voltage:  +4.0 VDC
  • Resolution: 1 mmHg
  • Bandwidth: tocotransducer DC to 0.5 Hz, strain gauge DC to 3 Hz
  • Zero temperature drift: < 0.1 mmHg/°C (electronics only).
  • Leakage current:  <10 µA at 120 VAC (transducer isolated).
Recorder
  • Uterine activity: chart width: 4 cm, scaling: 25 mmHg/cm (3.33 kPa/cm), range: 0-100 mmHg, resolution: 1 mmHg. 
  • Chart drive: speeds: 1, 2 and 3 cm/min; accuracy: ±1%.
  • Heart rate scale: domestic - chart width 7 cm, scaling 30 BPM/cm, range 30-240 BPM, resolution 1 BPM; international: chart width 8 cm, scaling 20 BPM/cm, range 50-210 BPM, resolution 1 BPM
Power Requirements
  • Voltage: 100/120/220/240 VAC selectable
  • Frequency: 50/60 Hz.
  • Power consumption:  100 Watts (max)
Physical Characteristics
  • 7" H x 16.5" W x 17" D (17.8 x 41.9 x 43.2 cm)
  • Weight:  approximately 34 lbs (15.5 kg) 


TOP^