englishespañol
printer friendly versionemail webpage link

DRE Axis 4 Portable Vital Signs Monitor

Add
Model Number:
66007
Condition:
  • Factory New
download sales sheetSales Sheet
More Resources
Manuals, Docs, Links
DRE Axis 4 Portable Vital Signs Monitor
DRE Axis 4 Portable Vital Signs Monitor

Product Description

DRE’s Light, Portable Monitor Measures Temp, NIBP, ECG and SpO2

At less than 1.3 kg (2.87 lbs), the DRE Axis 4 provides vital signs monitoring that is mobile and accurate. It has a crisp, easy-to-read monographic LCD screen that displays waveforms for ECG leads I, II and III. The Axis 4 also displays data for SpO2, heart rate, pulse rate, SYS, DIA, mean, temp, time, battery status, alarm status, contrast, beep and alarm sound status, and trends.

FEATURES

  • Mobile monitoring: Provides up to three hours of on-the-move battery operation (up to five hours of battery operation for ECG-only use)
  • Designed for desktop, IV pole mount and handheld use
  • Demo mode for staff training
  • Monitors adult, pediatric and neonatal patients
  • Features a tabular trend display

 

SPECIFICATIONS

Physical Dimensions

  • Weight: 2.87 pounds (1.3 kg) 
  • Size: 3.3" x 9.6" x 3.3" / 85mm x 245mm x 84 mm (H x W x D)  

Display

  • Screen Type: Monographic LCD
  • Resolution: 240 x 64 pixels
  • Waveform Type: ECG leads I,II,III
  • Indicator: AC power or battery operated indicators
  • Display Data: SpO2 , heart rate, pulse rate, SYS, DIA, mean, temp, time, battery status, alarm status, contrast, beep and alarm sound status, trend

Trends

  • Memory: 15 day at 10 second intervals
  • Display: Tabular

Interface

  • S/W Upgrade: RS-232

AC Power

  • Input Voltage: 100 ~ 240VAC, 50/60Hz
  • Output Voltage: 12VDC, 3.5A

Battery

  • Type: Li-ion
  • Battery Run Time: 3 hours (Up to 5 hours for ECG only)

Environmental Temperature

  • Operating: 5º ~ 40º C (41º ~ 104º F)
  • Storage: -20º ~ 60º C (-4º ~ 140º F)

Humidity

  • Operating: 15 ~ 95% , non-condensing
  • Storage: 10 ~ 95% , non-condensing

Alarms

  • Audible and Visual Alarms: SpO2, pulse rate, heart rate, SYS, DIA, mean, temp, low battery, ECG lead off , SpO2 fi nger off , system error
  • Silence: 2 minutes
  • Volume: 0 (off ) ~ 7

ECG Performance

  • Leads: 3-lead I, II, III
  • Display Sensitivity: 1.0 cm/mV ±10%
  • Gain Selection: x0.25, x0.5, x1
  • Frequency Response: 0.5 ~ 40H
  • Sweep Speed: 25 mm/sec, 50 mm/sec
  • Pacemaker Detection: Yes
  • Heart Rate Range: 30 ~ 300 bpm
  • Heart Rate Accuracy: ± 3% or ± 3 bpm (whichever is greater)

Pulse Oximetry Performance

  • Measurement Range
    Sp02: 0 to 100%
    Pulse Rate: 25 to 300 bpm
  • Accuracy
    Sp02: 70 to 100% ±2%, 0 to 69% - unspecified
    Pulse Rate: 25 to 250 bpm ±3 bpm

Temperature

  • Measurement Range: 15º to 45º C (59º to 113º F)
  • Accuracy: ± 0.1 C (25º to 45º C); ± 0.2 C (15º to 24º C)
  • Probe Type: Compatible with YSI 400 series

NIBP Performance

  • Technique: Oscillometric
  • Measurement Range - Large Patient
    SYS: 30 ~ 270 mmHg
    DIA: 10 ~ 220 mmHg
    Mean: 20 ~ 225 mmHg
  • Measurement Range - Medium Patient
    SYS: 30 ~ 235 mmHg
    DIA: 10 ~ 220 mmHg
    Mean: 20 ~ 225 mmHg
  • Measurement Range - Small Patient
    SYS: 30 ~ 135 mmHg
    DIA: 10 ~ 110 mmHg
    Mean: 20 ~ 125 mmHg
  • Cuff Pressure Range: 0 ~ 280 mmHg
  • Pressure Accuracy: ±2% or ±3 mmHg (whichever is greater)
  • Static: ±5 mmHg average error
  • Clinical: 8 mmHg standard deviation
  • Pulse Rate Measuring Range: 40 ~ 240 bpm
  • Inflation Time for Cuff: Less than 40 sec. (standard adult cuff )
  • Total Cycle Time: 20 to 45 seconds typical (dependent on heart rate and motion artifact)
  • Intervals for AUTO Measurement Time (min): 5, 10, 15, 30, 60, 90, 120
  • Overpressure Protection: Hardware and software double protections
  • Adult: 315±10 mmHg
  • Pediatric: 265±10 mmHg
  • Neonate: 155±10 mmHg