Specifications:
Noninvasive Blood Pressure
- Determination Time: Automatic and Manual Modes, 20 to 45 seconds typical (120 seconds max); STAT mode, 19 seconds typical at nominal blood pressure and heart rate
- Range: Systolic, 30-245 mmHg; Diastolic, 10-210 mmHg; MAP, 20-225 mm Hg; Heart rate, 40-200 bpm
- Alarm Limits: systolic - high, 75-245 mmHg; low, 30-150 mmHg; diastolic - high, 50-210 mmHg; low 10-120 mmHg; MAP - high, 70.225 mmHg; low, 20-120 mmHg
ECG
- Heart Rate Range: 10-250 bpm
- Connector: AAMI recommended, ECGC; MS3102A-14S-6S
- Leads: Selectable, I, II, III Lead off detection
- Band Width: 0.5-40 Hz
- Sweep Speed: 12.5, 25 and 50 mm per second
- Gain: 1mV to 8mV full scale, selectable in 5 steps
- QRS Detection Tone: Volume selectable from 0(off) to 15
- Alarm Limits: Heart rate-high, 80-250 bpm; low, 10-140 bpm
Oxygen Saturation:
- Accuracy: (+/- standard deviation) ±2% at 90-99%; ±3% at 75-89%; unspecified under 75%
- Saturation Range: 40-99%
- Pulse Rate: From 30-250 bpm
- Pulse Averaging: User selectable for 0 to 30 sec. in increments of 3. Preset at 6 seconds at power turn on.
- Saturation Pitch Indicator: Pitch changes with saturation. Volume selectable from 0(off) to 15.
- Alarm Limits: Saturation - high, 90-99%; low, 40-95%
Alarm Functions:
- System: Internal system test performed at power-up and during operation
- Volume: Adjustable from 1 to 15 on all parameters
- NIBP: Procedural alarm for overpressure, excess determination time, excess pump up time, no determination
- SpO2: Procedural alarms for low signal, probe disconnect, probe failure
Mechanical:
- Weight: 9.8 lbs
- Dimensions: 9.25"H x 720"W x 8.5D
- Display: Electroluminescent, 320 x 256 pixels; viewing area, 3.8"W x 3.0"H
- Power Requirements: 88-268 VAC; 47-63 Hz
- Fuse Requirements: 2.0 amps
- Leakage Current: <100µA