User-friendly touch-screen patient monitor
Measures ECG, SpO2 , NIBP (Suntech technology), heart rate, temperature and respiration
Also available with dual invasive blood pressures, EtCO2 , a battery and a built-in recorder
- 12.1” color TFT display with a touch-screen interface
- Multi-lead ECG monitoring
- Built-in battery (optional)
- NIBP list display
- Built-in thermal recorder
- Networkable with a central monitoring system via a wireless or wired network
- 48 hours trend data analysis
- Compact design convenient for mobile monitoring
- Multi-lingual settings
Physical Dimensions & Weight
- Main Unit - Dim. 555mm H x 335mm W x 235mm D
- Weight - 5kg
Application
- Neonatal, pediatric and adult patients
Operation Environment
- Power Source - External AC power or internal battery
- AC Power - 100-240VAC, 50/60Hz, <150VA
- Battery - Rechargeable Lead-Acid; Operating time under the normal condition (1 hour); Operating time after the first alarm of low battery (10 min.)
Temperature
- Working - 5°-40° C
- Storage - -20°-65° C
Performance Specifications
- Display - 12.1” color TFT, Touch Screen
- Resolution - 800 x 600 pixels
- Trace - 4, 6 or 8 waveforms ECG (I, II, III, aVR, aVL, aVF, V1-V6), PLETH, RESP, IBPx2, ETCO2
- Indicator - Alarm indicator; Power indicator QRS beep and alarm sound
- Trend Time - 48 hours
- Recorder - Built-in, thermal array, 2 channels
- Record - Width 48mm
- Recorder - Paper 50mm
- Record - Speed 25mm/s, 50mm/s
ECG
- Input - 5-lead ECG cable and standard AAMI line for connection
- Lead Choice - I, II, III, aVR, aVF, aVL, V
- Gain Choice - x0.5, x1.0, x2.0
- Frequency Characteristic - 0.05-40 Hz
- ECG Waveforms - 7 channels
- Penetration Voltage - 4000VAC 50/60Hz
- Sweep Speed - 12.5, 25 and 50 mm/s
- HR Display Range - 30-300bpm
- Accuracy - ±1% or ±1bpm, whichever is greater
- Alarm Limit Range Setting - Upper limit 80-400bpm; Lower limit 20-150bpm
RESP
- Measure Method - RA-LL impedance
- Range - 0-120 rpm
- Accuracy - + 3 rpm
- Alarm Upper-lower Limit Setting - Upper limit 6-120 rpm, Lower limit 3-120 rpm
- Sweep Speed - 12.5 and 25 mm/s
NIBP (Suntech)
- Measuring Technology - Automatic oscillating measurement
- Cuff Inflating - <30s-0-300 mmHg, standard adult cuff
- Measuring Period - AVE<40s
- Mode - Manual, Auto, STAT
- Measuring Interval in AUTO Mode - 2 min-4 hrs
- Pulse Rate Range - 30-250 (bpm)
- Measuring Range - Adult/Pediatric Mode, SYS 40-250 (mmHg), DIA 15-200 (mmHg); Neonatal Mode, SYS 40-135 (mmHg), DIA 15-100 (mmHg)
- Resolution - 1mmHg
- Accuracy - Pressure, Maximum Mean error ±5mmHg, Max. Standard deviation 8mmHg
- Overpressure Protection - Adult Mode 280(mmHg), Neonatal Mode 150 (mmHg)
- Alarm Limit Setting - SYS 50-240 mmHg, DIA 15-180 mmHg
TEMP
- Range - 25°-50° C
- Accuracy - + 0.2 °C 25.0°-34.9° C; + 0.1° C 35.0°-39.9° C; + 0.2° C 40.0°-44.9° C; 0.3° C 45.0°-50.0° C
- Display Resolution - 0.1° C
- Alarm Upper-lower Limit Setting - Upper limit 0-50° C, Lower limit 0-50° C
- Channel - 2 channels
- Alarm Limit Setting - 10-50° C
SPO2
- ASpO2 Anti-motion SpO2
- SpO2% Range 0-100%
- SpO2 Accuracy + 2% 70-100%, nonmotion; + 3% 70-100%, motion
- Pulse Rate Range 30-250 bpm
- Pulse Rate Accuracy ±2 bpm (non-motion); ±3 bpm (motion)
- Alarm Upper-lower Limit Setting: Upper limit 70-100%, Lower limit 70-100%
- SpO2 Probe - Red Light LED Wavelength 660nm±5nm, Infrared Light LED
- Wavelength - 940nm±10nm
IBP (Option)
- Measurement Range - -10-300mmHg
- Channel - 2 channels
- Pressure Transducer Sensitivity, 5μV/V/mmHg
- Impedance Range - 300-3000Ω
- Transducer Sites - ART, PA, CVP, RAP, LAP, ICP
- Resolution - 1mmHg
- Accuracy - ±1mmHg or ±2%, whichever is greater
- Alarm range - -10-300mmHg
EtCO2 (Option)
- Mode of Sampling - Sidestream
- Principle of Operation -Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
- CO2 Measurement Range - 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
- CO2 Calculation Method - BTPS (Body Temperature Pressure Saturated)
- CO2 Resolution - 0.1mmHg(0-69mmHg); 0.25mmHg(70-150mmHg)
- CO2 Accuracy - 0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading; 71-100 mmHg ± 8% of reading; 101-150 mmHg ± 10% of reading; Above 80 breath per minute ± 12% of reading
- Sampling Rate - 100Hz
- Respiration Rate - 2-150 bpm
- Respiration Rate Accuracy - ±1 breath
- Response Time - <3 seconds - includes transport time and rise time
- Inspired CO2 Measurement Range - 3-50 mmHg