The DRE Spectra AG5. Two monitors in
one.
DRE introduces the multi-purpose Spectra AG5. It combines all the
features of a multiparameter patient monitor with the added benefits of an
anesthetic agent monitor. It has a large, high definition (12.2” diagonally)
color flat panel display, and is available with the following configurations:
- Base unit
- Strip-chart recorder
- ECG
- SpO2
- NIBP
- CO2
- Temp standard
- IP 2/Temp
FEATURES
- Large, electroluminescent, high-definition, flat panel
display
- Up to 8-channel selectable waveform display
- Easy-to-follow menu
- Programmable audible and visual alarm limits
- Built-in recorder
- Up to 96 hours of trend information
- 5 minutes of real-time waveform history for onscreen
review
- Memory/Recall: digital readouts of up to 250 patient
readings
- Adult to neonate usage
- AC/DC operation
- 2-hour battery backup
- Lifetime software updates
- Warranty: 2 years parts and labor
- Impedence respiration
- Standard with ECG, EtCO2, N2O, oxygen saturation, NIBP,
temperature, respiration, and anesthetic agents.
- Optional FIO2 and two invasive pressures
SPECIFICATIONS
Mechanical Description
- Size: 12” (H) x 7” (D) x 14” (W) / (25.4 cm x 13.3 cm x
31.7 cm)
- Weight: 16 lbs/18lbs with full options
Power Requirements
- Operation: AC/DC
- Internal Battery: 2x6 V 3.3 AH sealed lead-acid
- Battery: 2 hours
- Battery Charge Time: 16-20 hours
- AC Mains Input: Universal switching power
supply
hospital grade plug, 100-250 V / 50-60Hz, 16 VDC / 2.8 A output, UL,
CSA, GS
General Features
- Freeze Buffer: 5 minutes
- Trends: 6 minutes to 96 hours
- Alarms: Audio & Visual, User adjustable upper and
lower alarm limits
- Alarm Frequency:
Low alarm - 2.0-2.4 KHz
High
alarm - 3.0-3.4 KHz
ECG
- CMRR: 100 dB @ 60 Hz
- Heart Rate Range: 30-240 bpm
- Heart Rate Averaging: 4 beat average
- Heart Rate Accuracy: ±5 bpm or 10%, whichever is
greater
- Response Time of Heart Rate Meter for a Step: 80-40 bpm
less than 7 sec
- Change in Heart Rate: 80-120 bpm less than 11 sec
- Pacemaker Pulse Amp: ±2 mV to ±700 mV
- Rejection Duration: 0.1 ms to 2.0 ms
- Input: 5-lead ECG/respiration patient cable with no
series resistance
- Lead Selection: I, II, III (standard
configuration)
- ECG Reference Signal: 1 mV software generated
- Patient Isolation: 5000 VRMS 60 seconds 60 Hz
- Leakage Current:
a) Internal to Ground: <0.01 mA
(<10mA)
b) Enclosure: <0.01 mA (<10mA)
c) Leads-Off Sensing
Current:
AC current <0.01 mA
(<100mA)
DC current <0.01 mA (<10mA)
- Frequency Response/Band Width: 0.5-40 Hz (±dB)
- Patient Drive Current: < 10 mA
- Maximum T Wave Rejection Capability: 120% of QRS
segment
- Defibrillator & ESIS Protected: Tested with 5
kV
- Recovery Time Following Defib.: Within 8 seconds
- Lead Fault Alarm: Audible, visual
- QRS Indicator:
Audible: user volume
control
Asystole: audible and visual
- Gain Selection: 0.2 mV, 0.5 mV, 1 mV, 2 mV, 5 mV
- Aspect Ratio: 0.24, 0.4, 0.6, ±0.8 sec/mV
- ECG Input Impedance: >2.5 M Ohm
ETCO2
- Type: Side stream, non dispersive infrared
- Method: CO2 sample line with nasal cannula or
endotracheal tube connector
Invasive Pressure
- Channels: Two
- Measurement Range: -30 to +300 mmHg (-4.0 to +40.0
kPA)
- Gain Accuracy: ±1% or ±2 mmHg, whichever is greater
(±0.27 kPA)
- Analysis Values:
Max., min., and mean
pressure
Pulse rate from arterial waveforms
- Heart Rate Range: 10-250 bpm
- Heart Rate Accuracy: ±1% or ±5 bpm
- Transducer Sensitivity: 5mV/V/mmHg
- Transducer Excitation: 5V DC
- Transducer Zero Adjust: ± 150 mmHg (± 19.99 kPa)
- Display:
Digital waveform display with semiautomatic
and manual pressure range scaling.
Trend display of pressure
waveforms
Under Selectable Pressure Channel Label
IPI (2), ART1 (2), AO1
(2), RA1 (2), FA1 (2), CVP1 (2), PA (2)
Non-Invasive Blood Pressure
- Method: Automatic oscillometric
- Parameters Measured: Systolic, diastolic, mean arterial
pressure, pulse
- Scale: mmHg or kPa
- Operating Modes: Manual, Automatic, Stat
- Repeat Cycles: 10-50 sec., 1-99 minutes
- Rapid Cycle Update: 1-4 minutes (STAT mode)
- Measurement Range:
Systolic:
Adult/pediatric - 30-250 mmHg (4.0-33.3kPa)
Neonate - 20-160
mmHg (2.7-21.3kPa)
Diastolic:
Adult/pediatric - 10-180 mmHg
(1.3-24.0kPa)
Neonate - 10-140 mmHg (1.3-18.7kPa)
- Measurement Time:
Typical - 50 sec. Max. 120
sec.
Typical Stat - 30 sec.
- Cuff Inflation Rate: Not greater than 40-50 mmHg/sec
(5.33-6.66 kPa/sec)
- Cuff Inflation Pressure: 30 mmHg above last systolic
(4.0kPa)
- Cuff Pressure Range:
Adult/pediatric - 0-250 mmHg
(0-33.3 kPa)
Neonate - 0-140 mmHG (0-18.7 kPa)
- Initial Cuff Inflation:
Adult/pediatric - 170 ± 10
mmHg (22.7 ±1.3 kPa)
Neonate - 120 ± 10 mmHg (16 ±1.3 kPa)
- Auto Deflate Pressure:
Adult/pediatric - 280 ±5 mmHg
(36.7-38 kPa)
Neonate - 235 ±5mmHg (30.7-32.0 kPa)
- NIBP Display Accuracy
- NIBP Alarm Limits:
Systolic:
Upper -
0-255 mmHg (0-34.0 kPa)
Lower - 0-254 mmHg (0-33.9
kPa)
Diastolic:
Upper - 0-255 mmHg (0-34.0
kPa)
Lower - 0-254 mmHg (0-33.9 kPa)
Mean:
Upper - 0-255 mmHg (0-34.0 kPa)
Lower - 0-254 mmHg (0-33.9
kPa)
Temperature
- Temp Scale: °F or ºC
- Temp Range: 82.4-109.8ºF (28.0-43.2ºC)
- Temp Accuracy: ±0.2ºF (±0.1ºC)
- DTemp Range: 0-27ºF (0-15ºC)
- Temp Alarm Limits:
Upper - 83.0-110.0ºF (28.3-43.3ºC)
Lower - 83.0-109.9ºF (28.3-43.2ºC)
Recorder
- Type: Built-in multi-channel thermal array recorder
- Dot Density: 203 DPI
- Paper Width: 58 mm
- Printing Width: 48 mm
- Speed: 25 mm/sec