BCI 3303 Hand-held Pulse Oximeter
Cost-effective, portable, continuous, or spot check pulse oximetry monitoring system with alarms and an internal rechargeable battery. Clinician, Home-Use and Sleep Study modes make it ideal for use in the hospital or clinical environment; during emergency air or land transport; or for in-home use.
The Smiths Medical PM, Inc. advantage:
TECHNICAL SPECIFICATIONS
Displays, Indicators & Keys:
SpO2: LED numeric display
0.43 inch (11 mm) high
Pulse Rate: LED numeric display
0.375 inch (9.5 mm) high
Pulse Strength: Logarithmically scaled 8-segment LED bar graph
Low Battery: LED flashes when approximately 30 minutes of battery life remains
Alarm Silence: LED flashes or lights to remind the user that audio alarms are disabled
Sensor: LED flashes to alert the operator to check the SpO2 sensor placement
SpO2:
Range: 0-100%
Accuracy: ±2% at 70-100%; ±3% at 50-69%
Averaging: 4, 8, or 16 pulse beat average
Alarm Range: High - 50%-100% & OFF (1% steps)
Low - OFF & 50%-99% (1% steps)
Pulse Rate:
Range: 30-254 bpm
Accuracy: ±2% at 30-254 bpm
Resolution: 1 bpm
Alarm Range: High - 5-250 bpm & OFF (5 bpm steps)
Low - OFF & 5-250 bpm (5 bpm steps)
Memory:
Spot checks and trends can be stored for up to 99 patients and 24 hours of run time.
Auxiliary Serial Printer/Computer Output:
Data saved every thirty seconds can be printed or downloaded to a PC (trend). Data saved every four seconds in sleep mode can be output and processed by commercially available sleep study software.
Physical:
Dimensions: 7.25” H x 3.3” W x 1.85” D
18.4 cm x 8.4 cm x 4.7 cm
Weight: 19 oz. (539 grams)
Power:
AC Power: 105-125V ~ 60Hz
Optional: 207-253V ~ 50Hz, 90-110V ~ 50Hz
Battery Standard: NiMH (Nickel Metal Hydride)
Battery Life: Internal rechargeable, not user replaceable. Fully charges in about 6 hours. Approximately 24 hours of continuous use.
Environment:
Operating Temperature: -4°F to 131°F (-20°C to 55°C)
Storage Temperature: -40°F to 167°F (-40°C to 75°C)
Relative Humidity (non-condensing): 10-95% (storage)
15-95% (operating)
Mechanical Durability Test Results:
Vibration Testing: In compliance with IEC 68-2-6, IEC 68-2-34
Shock Testing: In compliance with IEC 68-2-27
Drop Testing: Exceeded UL 544 testing recommendations
Standards: EN60601-1, EN60601-1-2, CSA Std. C22.2 No. 125, UL 544
Regulatory Compliance: 510(k): K945754
ISO 9001/ISO 13485
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE
Please read the Instructions for Use supplied with the product for detailed instructions, warnings, and cautions.