DRE Medical Equipment
englishespañol
phone: 1-800-462-8195DRE Facebook PageDRE LinkedIn PageDRE Twitter Profile

Invivo MDE Prism

Condition:
  • Refurbished
Invivo MDE Prism

Product Description

 MDE Escort Prism SE model 20414-103 patient vital signs monitor.

MDE Escort Prism SE model 20414-103 vital signs patient monitor with ecg/resp, Nellcor SPO2, NIBP, temp, turbo-temp, & printer; 900 MHz transceiver available.

Large color screen. The portable MDE Escort Prism SE has a battery and is easily transportable for any situation. This Escort Prism SE is the newer style of the popular Escort family, with BUILT-IN parameters including ecg/resp, SPO2, NIBP, temp, as well as Turbo-temp and a printer. The large color monitor screen is great and can be adjusted for optimal viewing. This unit is also available with the 900 MHz transceiver, model 20011V, which is to be used with the Vision Central Station system.

Display Type: CRT non-fade P31. Size: 6” (15.2 cm) diagonal. Number of Traces: 3. Trace Length: 3.2 sec @ 25 mm/sec
Front Panel Key Functions: Freeze, Record, Alarm Suspend, Alarm Set Up, Page Home, Next Page, 8 Softkeys, On/Standby.

ECG—
Range: 10 to 300 BPM
Accuracy: ± 2% or ± 2 BPM
Sweep Speed: 12.5, 25, 50 mm/sec selectable
High Alarm Limit Range: 50 to 250 BPM in 5 BPM steps – ADULT. 60 to 255 BPM in 5 BPM steps – PED/NEO.
Low Alarm Limit Range: 20 to 150 BPM in 5 BPM steps – ADULT. 20 to 150 BPM in 5 BPM steps – PED/NEO.
Bandwidth: 0.5 to 40 Hz (monitor); 0.05 to 100 Hz (diagnostic)
Leads: 3 lead: I, II, or III selectable. 5 lead, dual vector: I, II, or III, and V selectable
Sensitivity: 0.25 to 3 mV/cm selectable (12 levels)
Pacer Rejection: 0.1 to 2 msec; 2 to 700 mV
Heart Rate Source: ECG, Pleth, IBP
Defibrillation Tolerance: 400 joule with 50 ohm series lead impedance
Connector: Standard 6 pin MS
Isolation: Full electrical isolation

Respiration—
Range: 4 to 200 BPM
Accuracy: ± 2% or ± 2 BPM
Lead: RA to LA
Sensitivity: 0.25 to 3 ohm/cm selectable
High Alarm Limit Range: 30 to 150, OFF – Adult. 50 to 200, OFF – PED/NEO
Low Alarm Limit Range: 5 to 50, OFF – Adult. 5 to 80, OFF – PED/NEO
Apnea Delay Alarms: 10, 15, 20, 30, 40 seconds, and OFF – ADULT. 10, 15, 20 seconds, and OFF – PED/NEO
Sweep Speed: 1.56, 6.25, 12.5, 25 mm/sec selectable
Respiration Rate Source: ETCO2 or Leads
CVA Rejection: Rejects most cardiovascular coincidence.