Refurbished - Infrasonics Adult Star 2000 Ventilator

Condition:
  • Refurbished
Infrasonics Adult Star 2000 Ventilator

Product Description

FEATURES

  • Adult ventilator with graphics display and compressor.
  • Fully comprehensive, microprocessor controlled and monitored adult/pediatric volume cycled, pressure limited ventilator.
  • A single knob adjusts all ventilator settings “on screen”.
  • Graphic monitoring provides real time pressure, flow and volume waveforms.

TECHNICAL SPECIFICATIONS

  • Modes of Ventilation
    - Ventilation Type: Volume cycled (pressure control, time cycled or I:E cycled optional)
    - Modes: assist/control, SIMV, CPAP and apnea ventilation
     
  • Controls and Displays
    - Tidal Volume (Compliance Compensated): 100 to 2500 ml (50 to 2500 ml - pediatric option)
    - Machine Rate / Apnea Rate: 0.5 to 80 BPM (0.5 to 120 BPM - pediatric option)
    - Peak Flow (BTPS Compensated): 10 to 120 LPM volume cycled 10 to 160 LPM spontaneous, PPS and PCV
    - Oxygen Percent: 21 to 100% (integral blender). Educated button provides 100% for 3 minutes.
    - PEEP / CPAP: 0 to 30 cmH2O (50 cmH2O optional)
    - Sensitivity: -0.5 to -20 cmH2O
    - Inspiratory Pause: 0 to 2 seconds
    - Waveform: Square, sinusoidal, accelerating or decelerating
    - Positive Pressure Support: 0 to 70 cmH2O (above PEEP)
    - Sigh Volume: 100 to 2500 ml
    - Sigh Rate: 0 to 20 sph, multiple of 1 to 3
    - Manual Breath: Single breath
    - Manual Sigh: Single Sigh
     
  • Alarms and Indicators
    - High Inspiratory Pressure Alarm: 10 to 120 cmH2O
    - Low Inspiratory Pressure Alarm: 3 to 60 cmH2O
    - High and Low Mechanical Volume Alarm: 0 to 2500 ml
    - Low Spontaneous Volume Alarm: 0 to 2500 ml
    - Low Minute Volume Alarm: 0 to 60 LPM
    - Low PEEP / CPAP Alarm: 0 to 30 cmH2O (50 cmH20 optional)
    - High Respiratory Rate Alarm: 0 to 90 BPM (0-130 BPM - pediatric option)
    - Apnea Alarm: 5 to 60 seconds
    - Low Pressure Oxygen and Air Alarms: <26 psig ±3 psig
    - Low Internal Battery Alarm: Last 4 minutes of battery operation
    - High PEEP / CPAP: 10 cmH2O >PEEP/CPAP for 3 seconds, 6 to 10 cmH2O >PEEP/CPAP for 5 seconds
    - Airway Leak Alarm: 6 LPM
    - Insufficient Inspiration Alarm: > 240 milliseconds
    - Insufficient Expiratory Alarm: > 200 milliseconds
    - Exhalation Leak: Exhaled flow > 4 LPM and > 10% of delivered flow
    - Obstructed Tube: Tube or prox pressure line occluded
    - Alarm Silence: 60 seconds
     
  • Monitored Parameters
    - Exhaled Tidal Volume: 0 to 2500 ml
    - Minute Volume: 0 to 99.9 LPM
    - Spontaneous Minute Volume: 0 to 99.9 LPM
    - Breath Type: Assisted, controlled, spontaneous or sigh
    - Peak and Mean Airway Pressures: 0 to 120 cmH2O
    - PEEP / CPAP Pressure: 0 to 30 cmH2O (50 cmH2O optional)
    - Plateau Pressure: 0 to 120 cmH2O
    - Breath Rate: 0 to 120 BPM
    - I:E Ratio: 99.9:1 to 1:99.9
    - Inspiratory Time: 0 to 99.9 seconds
    - Proximal Airway Pressure Bar Graph: -20 to 120 cmH2O
    - Graphic Monitor (Waveform Display):
    Pressure: -20 to 120 cmH2O
    Flow: -200 to 160 LPM
    Volume: -100 to 3000 ml
    Time: 6-12-24 seconds freeze
     
  • Exhalation Isolation System
    - Heated Exhalation Bacteria Filer: Filtration ability 99.99% of particles > 0.3 microns
     
  • General
    - Integral Nebulizer: Provides intermittent 8 LPM @ 10 psig nominal output. Automatically turns off after 30 minutes
    - Integral Compressor: Low noise, 60 dB
    - Alarm Loudness: Adjustable, 64 to 82 dB
    - Analog Pressure Output: 10 mV/cmH2O
     
  • Electrical Power
    Nominal: 115 VAC, 60 Hz
    - Range: 108 to 132 VAC, 60 Hz ±3 Hz
    - Power Consumption: Ventilator 1.5 amp, ventilator and nebulizer 3.0 amp, ventilator, nebulizer and compressor 7.0 amp.
    Nominal: 230 VAC, 50/60 Hz
    - Range: 207 to 242 VAC, 50/60 Hz ±3 Hz
    - Power Consumption: Ventilator 1.0 amp, ventilator and nebulizer 1.7 amp, ventilator nebulizer and compressor 3.2 amp
    Nominal: 100 VAC, 50/60 Hz
    - Range: 90 to 105 VAC, 50/60 Hz ±2 Hz
    - Power Consumption: Ventilator 1.7 amp, ventilator and nebulizer 3.2 amp, ventilator, nebulizer and compressor 7.5 amp
     
  • Pneumatic Power
    - Air and Oxygen: 30 to 90 psig (2.1 to 6.2 bars) @ 160 LPM
     
  • Physical Dimensions and Weight
    (with computer and base enclosed)
    - Dimensions: 54.5" height x 22.5" wide x 22.5" deep A(138 x 57 x 57 cm)
    - Weight: 185 lbs (84 kg)

 