Product Description
Superior flexibility and performance with a premium flat panel display
The DRE Waveline Pro displays as many as eight colorful waveforms on a crisp 15” high resolution display. It is lightweight, portable and measures multiple parameters, including ECG (3 or 5 lead) with ST segment monitoring, resp., dual temp., SpO2 and NIBP. Configurations are available that measure capnography, dual channel IBP, and anesthetic agents (five agents, N2O, FiO2, EtCO2), and that have a built-in thermal printer.
FEATURES
- Displays as many as eight waveforms
- Available with automatic agent ID
- Multi-lead simultaneous ECG monitoring
- Automatically set alarm limits
- Graphical and tabular trending
- Color-coded alarms
- Patient data entry
- Battery backup (two removable batteries)
- Networking capability
- Large (15”), high resolution flat panel display
- Available with a built-in thermal printer
SPECIFICATIONS
Physical Dimensions & Weight
- Base Unit: 12.1”/309 mm (H) x 14.4”/365 mm (W) x 6.3”/159 mm (D)
- Weight: 17.6 lbs/8.0 kgs (including anesthetic agents module and batteries)
Application
- Neonatal, pediatric and adult patients
Performance Specifications
- Display: 15” color TFT, Resolution 1024 x 768
- Trace: 8 waveforms
- Indicators:
Alarm indicator
Power indicator
QRS beep and alarm sound - Trend time: 1 – 48 hours
- Recorder:
Built-in, thermal array, 2 channels
Record width: 1.9”/48 mm
Recorder paper: 2”/50 mm
Record speed: 25 mm/s, 50 mm/s
ECG
- Input: 5-lead ECG cable and standard AAMI line for connection
- Lead Choice: I, II, III, aVR, aVF, aVL, V, TEST
- Gain Choice : x0.5, x1, x2, x4
- Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
- ECG Waveforms: 7 channels
- Penetration Voltage: 4000VAC 50/60Hz
- Sweep Speed: 12.5, 25 and 50 mm/s
- HR Display Range: 30 – 300bpm
- Accuracy: ±1 bpm or ±1%, whichever is greater
- Alarm Limit Range Setting: Upper limit 100 – 200 bpm; Lower limit 30 – 100 bpm
RESP
- Measure Method: RA-LL impedance
- Range: 0 – 120 rpm
- Accuracy: ±3 rpm
- Alarm Limit Setting: Upper limit 6 – 120 rpm: Lower limit 3 – 120 rpm
- Sweep Speed: 12.5 and 25 mm/s
TEMP
- Range: 25° – 50° (C)
- Accuracy:
± 0.2° C (25.0° – 34.9° C)
± 0.1° C (35.0° – 39.9° C)
± 0.2° C (40.0° – 44.9° C)
± 0.3° C (45.0° – 50.0° C) - Display Resolution: 0.1° C
- Alarm Limit Setting: Upper limit 0° – 50° C, lower limit 0° – 50° C
- Channel: 2 channels
SpO2
- ASpO2: Anti-motion SpO2
- SpO2% Range: 0 – 100%
- SpO2 Accuracy: ±2% (70 – 100%, non-motion) ±3% (70 – 100%, motion)
- Pulse Rate Range: 30 – 250 bpm
- Pulse Rate Accuracy: ±2 bpm (non-motion) ±3 bpm (motion)
- Alarm Limit Setting: Upper limit 70 – 100%, lower limit 70 – 100%
- SpO2 Probe:
Red light LED wavelength – 660nm±5nm
Infrared light LED wavelength – 940nm±10nm
NIBP
- Measuring Technology: Automatic oscillating measurement
- Cuff Inflating: <30s (0 – 300 mmHg, standard adult cuff)
- Measuring Period: AVE<40s
- Mode: Manual, Auto
- Measuring Interval in AUTO Mode: 2 min – 4 hrs
- Pulse Rate Range: 30 – 250 (bpm)
- Measuring Range:
Adult/Pediatric Mode
- SYS: 40 – 250 (mmHg) DIA :15 – 200 (mmHg)
Neonatal Mode
- SYS: 40 – 135 (mmHg) DIA : 15 – 100 (mmHg) - Resolution: 1mmHg
- Accuracy:
Maximum Mean error: ±5mmHg
Maximum Standard deviation: 8mmHg - Overpressure Protection:
Adult Mode: 300(mmHg)
Neonatal Mode: 160 (mmHg) - Alarm Limit Setting:
- SYS: 50 – 240 mmHg
- DIA: 15 – 180 mmHg
IBP
- Measurement Range: -50 – 300mmHg
- Channel: 2 channels
- Pressure Transducer: Sensitivity, 5μV/V/mmHg
- Impedance Range: 300 – 3000Ω
- Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
- Unit: mmHg/kPa selectable
- Resolution: 1 mmHg
- Accuracy: ±1 mmHg or ±2%, whichever is greater
- Alarm Range: -10 – 300 mmHg
EtCO2 (Oridion)
- CO2 Measurement Range: 0 – 99 mmHg
- Accuracy:
±2 mmHg (0 – 38 mmHg)
39-99 mmHg ±5% of reading + 0.08% for every 1 mmHg (above 38 mmHg) - Sampling Rate: 50 ml/min
- Initialization Time: 30 seconds (typical), reaches ±5% steady-state accuracy within 3 minutes
- Respiration Rate: 0 – 150 breaths/min
- Mode: Adult, neonate
Anesthetic Agents
- Method: Infrared absorption
- Gas Sorts: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, O2 (optional Automatic Agent ID)
- Measurement Range:
Halothane, Isoflurane: 0 – 8.5%
Enflurane, Sevoflurane: 0 – 10%
Desflurane: 0 – 20%
CO2: 0 – 10%
N2O: 0 – 100%
O2: 0 – 100% - Bias:
Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane: ±(0.15 Vol% + 15% rel.)
CO2: ±(0.5 Vol% + 12% rel.)
N2O: ± (2 Vol% + 8% rel.) - O2: ±3 Vol%
Networking
- Industry standard 802.11b/g wireless network
Power
- Source: External AC power or internal battery
- AC Power: 100 – 240VAC, 50/60Hz, 150VA
- Battery: Built-in and rechargeable
- Charge Time: 4 hours
Environmental Specifications
- Temperature:
Operating: 5° – 40° C
Storage: -20° – 65° C - Humidity range:
Operating: ≤80 %
Storage: ≤80 %