The MINI-TORR PLUS NON-INVASIVE BLOOD PRESSURE MONITOR provides a high quality reliable platform for your NIBP monitoring needs. Its combination of reliability and flexibility makes the Mini-Torr Plus® a economical starting point for your long-range NIBP monitoring needs.
SPECIFICATIONS:
Display:
Pulse Rate 3-digit LED display, 0.30 inches (7.62 mm) high
Pulse Strength Logarithmically scaled 10-segment
LED bar graph
NIBP 3-digit LED display
Systolic & Diastolic Pressures
0.40 inches (10.16 mm) high
Mean Arterial Pressure
0.30 inches (7.62 mm) high
Temperature 3-digit LED display, 0.40 inches (10.16 mm) high
NIBP:
Adult: 20 to 250 mmHg
Neonate: 20 to 135 mmHg
Cuff Pressure ±3 mmHg or 2%
Determination 30 to 50 seconds typical
Time (max. 120 seconds in adult mode, 90 seconds in neonate mode)
Interval Times 1, 2, 2.5, 5, 10, 15, 20, 30, 45, 60, 90 mins
Calibration Factory calibrated
Inflation Pressure Adult: 50 to 300 mmHg
Neonate: 50 to 150 mmHg
Default Inflation Adult: 175 mmHg
Pressure Neonate: 100 mmHg
Alarm Range
Adult High: 15 to 285 mmHg & OFF
Low: 10 to 280 mmHg & OFF
Neonate High: 15 to 135 mmHg & OFF , Low: 10 to 130 mmHg & OFF
Alarm Default
Adult Systolic: High - 200 mmHg , Low - 90 mmHg
Diastolic: High - 105 mmHg , Low - 40 mmHg
Mean Arterial: High - 140 mmHg , Low - 50 mmHg
Pulse Rate:
Range 30 to 180 bpm (NIBP) (adult)
40 to 180 bpm (NIBP) (neonate)
30 to 254 bpm (SpO2)
Accuracy ±2% or 2 bpm (whichever is greater)
Alarm Range
Adult & Neonate High: 30 to 250 bpm & OFF (1bpm steps) , Low: 30 to 250 bpm & OFF (1bpm steps)
Alarm Default Adult: High: 150 bpm , Low: 45 bpm
Neonate: High: 150 bpm , Low: 50 bpm
SpO2 (Optional):
Range 0 to 100 SpO2 (functional)
Accuracy ±2 counts at 70 to 100%
<70% unspecified
Averaging 8 beats
Pulse Tone Volume adjustable or OFF
Alarm Default Adult: High: 150 bpm , Low: 45 bpm
Neonate: High: 150 bpm , Low: 50 bpm
Temperature (Optional):
Range 84°F to 108°F (28.9°C to 42.2°C)
Accuracy ±2°F, ASTM E1112-86 (±0.1°C)
±3 counts at 50 to 69%
Calibration Factory calibrated
Dimensions:
Base Unit 3.24” H x 8.5” W x 5.5” D
With Temp 3.24” H x 10” W x 5.5” D
Weight:
Base Unit 3.5 pounds (1.6 kg)
With Temperature 4.2 pounds (1.9 kg)
Battery Life
Fully charged continuous-life 6 hours approximate, charge time 4 hours
Printer Output:
RS232C
Data Format: 19200 baud
Integrated Printer (Optional):
Text only, no graphics. Patient data log or trend tables.