Product Description
Effective Capnography Monitoring.
The DRE Echo CO2 provides cost-effective SPO2 and CO2 monitoring with
accuracy and durability. The Echo CO2 ensure effective capnography monitoring for intubated and non-intubated patients for continuous long-term monitoring. It is tailored for mechanically ventilated and non-intubated patients.
FEATURES
- 5.7 inch high resolution display for easy reading
- Lightweight, portable design and user-friendly interface for easy operation
- Flexible configurations to meet different clinical needs
- Standard SpO2
- PR measurement
- Respironics Loflo sidestream or CAPNOSTAT 5 EtCO2 mainstream measurement
- Real-time parameters measurement display with trend table for easy reviewing
- Built-in rechargeable Lithium-ion battery for 10 hours continuous working
- Powerful storage capacity
- Bi-directional communications with central station by wired or wireless network
- Nurse call
- Suitable for adult, pediatric and neonatal patient
SPECIFICATIONS
- Size: 200.8mm (L) x 241mm (H) x 189 mm (D)
- Weight: 3 kg
- Display: 5.7 inches
- Color TFT resolution: 640X480
- Power Supply: 100-240 VAC, 50/60HZ
- Pmax=70VA FUSE T 1.6AL
- Battery Type: Lithium-ion
- Voltage: 14.8 V DC Capacitance: 4,400 mAh
- Working period: Color TFT 480min
- Rechargeable period: < 360min
- Thermal Recorder: (Optional)
- Record Width: 48 mm
- Paper Speed: 25mm/s
SpO2 (Standard)
- Measuring Range: 0 ~ 100 %
- Alarm Range: 0 ~ 100 %
- Resolution: 1 %
Accuracy
Adult (including Pediatric)
± 2% (70%~100% SpO2)
Undefined 0~70% SpO2
Neonate
± 3% (70%~100% SpO2)
Undefined (0~70% SpO2)
Pulse Rate
Measuring and Alarm Range: 30 ~ 300bpm
Resolution: 1bpm
Accuracy: ± 3bpm
Data update period: 2s
ISO 9919
SpO2 (Option, by Nellcor OxiMax)
- Measuring Range: 1 ~ 100 %
- Alarm Range: 1 ~ 100 %
- Resolution: 1 %
Accuracy:
Adult (including pediatric)
±2% (70%~100% SpO2)
Undefined: (0~70% SpO2)
Neonate:
±3% (70%~100% SpO2)
Undefined: (0~70% SpO2)
Pulse Rate:
Measuring and Alarm Range: 20~300 bpm
Resolution: 1 bpm
Accuracy: 3 bpm
Respironics CO2
Mainstream
- CO2 Measurement Range
0 to 150 mm Hg,0 to 19.7%,
0 to 20 kPa (at 760 mm Hg)
- Rise Time
Less than 60 ms Adult/Infant
Reusable or Single Patient Use
Airway Adapter
- CO2 Resolution
0.1 mm Hg -- 0 to 69 mm Hg
0.25 mm Hg -- 70 to 150 mm Hg
- CO2 Accuracy
0-40 mm Hg -- ±2 mmHg
41 - 70 mm Hg -- ± 5% of reading
71 - 100 mm Hg -- ± 8% of reading
101 - 150 mm Hg -- ±10% of reading
Above 80 bpm -- ±12% of reading
- CO2 Stability
Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
Long Term Drift: Accuracy specification will be maintained over a 120-hour period
- Respiration Rate Accuracy
±1 breath
- Calibration
No routine user calibration required. An airway adapter zero is required when
changing to a different style of airway adapter
- Water Resistance
IPX4-Splash-proof (sensor head only)
- Shock Impact
EN60068-2-6 Sinusoidal Vibration
EN60068-2-27 Shock
EN60068-2-64 Random Vibration
Able to withstand repeated 6 foot drops
onto tiled floor while operating
Sidestream
- Sample Rate: 50 m/L per minute
- CO2 Measurement Range:
0 - 150 mm Hg, 0 to 20 kPa
(at 760 mm Hg). Barometric Pressure supplied by Host
- CO2 Resolution:
0.1 mm Hg 0 to 69 mm Hg
0.25 mm Hg 70 to 150 mm Hg
- CO2 Accuracy
0 - 40 mm Hg ±2 mm Hg
41 - 70 mm Hg ±5% of reading
71 - 100 mm Hg ±8% of reading
101 - 150 mm Hg ±10% of reading
Above 80 bpm ±12% of reading
- CO2 Stability
Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
Long Term Drift: Accuracy specification will be maintained over a 120-hour period
- Respiratory Rate Range: 2 to 150 bpm
- Respiratory Rate Accuracy: ±1 breath
- Sample Cell/Filter
Proprietary single patient use sample cell and inline filter are integrated with the sample
line which eliminates contamination of the internal system.
- Nasal Sampling Kits for Non-intubated Patients
Adult, pediatric and infant nasal CO2 sampling, nasal CO2 sampling and O2 delivery
Adult and pediatric nasal/oral CO2sampling, nasal/oral CO2 sampling and O2 delivery
- On-Airway Adapter KITS for Intubated Patients
Adult/Pediatric with and without dehumidification tubing
Pediatric/Infant, low dead space, with and without dehumidification tubing
Taper meets ISO 5356-1
- Sample Kit Hours of Use
Nasal Cannula (all styles) — up to 12 hours
On-Airway Adapter Kits without dehumidification tubing — up to 12 hours
- Sample Cell Detection
Insertion automatically turns sampling pump on. Removal automatically turns sampling pump off.
- Water Resistance IPX4 -Splash-proof (when sample cell is inserted in sample cell receptacle)
- Shock ImpactIEC TR 60721-4-7 Class 7M3
(designed to withstand environments subject to significant vibrations or high shock levels)
EN60068-2-27 Shock
EN60068-2-64 Random vibration
Specifications subject to change without notice.