A single knob adjusts all ventilator settings “on screen”.
Graphic monitoring provides real time pressure, flow and volume waveforms.
TECHNICAL SPECIFICATIONS
Modes of Ventilation
- Ventilation Type: Volume cycled (pressure control, time cycled or I:E cycled optional)
- Modes: assist/control, SIMV, CPAP and apnea ventilation
Controls and Displays
- Tidal Volume (Compliance Compensated): 100 to 2500 ml (50 to 2500 ml - pediatric option)
- Machine Rate / Apnea Rate: 0.5 to 80 BPM (0.5 to 120 BPM - pediatric option)
- Peak Flow (BTPS Compensated): 10 to 120 LPM volume cycled 10 to 160 LPM spontaneous, PPS and PCV
- Oxygen Percent: 21 to 100% (integral blender). Educated button provides 100% for 3 minutes.
- PEEP / CPAP: 0 to 30 cmH2O (50 cmH2O optional)
- Sensitivity: -0.5 to -20 cmH2O
- Inspiratory Pause: 0 to 2 seconds
- Waveform: Square, sinusoidal, accelerating or decelerating
- Positive Pressure Support: 0 to 70 cmH2O (above PEEP)
- Sigh Volume: 100 to 2500 ml
- Sigh Rate: 0 to 20 sph, multiple of 1 to 3
- Manual Breath: Single breath
- Manual Sigh: Single Sigh
Alarms and Indicators
- High Inspiratory Pressure Alarm: 10 to 120 cmH2O
- Low Inspiratory Pressure Alarm: 3 to 60 cmH2O
- High and Low Mechanical Volume Alarm: 0 to 2500 ml
- Low Spontaneous Volume Alarm: 0 to 2500 ml
- Low Minute Volume Alarm: 0 to 60 LPM
- Low PEEP / CPAP Alarm: 0 to 30 cmH2O (50 cmH20 optional)
- High Respiratory Rate Alarm: 0 to 90 BPM (0-130 BPM - pediatric option)
- Apnea Alarm: 5 to 60 seconds
- Low Pressure Oxygen and Air Alarms: <26 psig ±3 psig
- Low Internal Battery Alarm: Last 4 minutes of battery operation
- High PEEP / CPAP: 10 cmH2O >PEEP/CPAP for 3 seconds, 6 to 10 cmH2O >PEEP/CPAP for 5 seconds
- Airway Leak Alarm: 6 LPM
- Insufficient Inspiration Alarm: > 240 milliseconds
- Insufficient Expiratory Alarm: > 200 milliseconds
- Exhalation Leak: Exhaled flow > 4 LPM and > 10% of delivered flow
- Obstructed Tube: Tube or prox pressure line occluded
- Alarm Silence: 60 seconds
Monitored Parameters
- Exhaled Tidal Volume: 0 to 2500 ml
- Minute Volume: 0 to 99.9 LPM
- Spontaneous Minute Volume: 0 to 99.9 LPM
- Breath Type: Assisted, controlled, spontaneous or sigh
- Peak and Mean Airway Pressures: 0 to 120 cmH2O
- PEEP / CPAP Pressure: 0 to 30 cmH2O (50 cmH2O optional)
- Plateau Pressure: 0 to 120 cmH2O
- Breath Rate: 0 to 120 BPM
- I:E Ratio: 99.9:1 to 1:99.9
- Inspiratory Time: 0 to 99.9 seconds
- Proximal Airway Pressure Bar Graph: -20 to 120 cmH2O
- Graphic Monitor (Waveform Display):
Pressure: -20 to 120 cmH2O
Flow: -200 to 160 LPM
Volume: -100 to 3000 ml
Time: 6-12-24 seconds freeze
Exhalation Isolation System
- Heated Exhalation Bacteria Filer: Filtration ability 99.99% of particles > 0.3 microns
General
- Integral Nebulizer: Provides intermittent 8 LPM @ 10 psig nominal output. Automatically turns off after 30 minutes
- Integral Compressor: Low noise, 60 dB
- Alarm Loudness: Adjustable, 64 to 82 dB
- Analog Pressure Output: 10 mV/cmH2O
Electrical Power
Nominal: 115 VAC, 60 Hz
- Range: 108 to 132 VAC, 60 Hz ±3 Hz
- Power Consumption: Ventilator 1.5 amp, ventilator and nebulizer 3.0 amp, ventilator, nebulizer and compressor 7.0 amp.
Nominal: 230 VAC, 50/60 Hz
- Range: 207 to 242 VAC, 50/60 Hz ±3 Hz
- Power Consumption: Ventilator 1.0 amp, ventilator and nebulizer 1.7 amp, ventilator nebulizer and compressor 3.2 amp
Nominal: 100 VAC, 50/60 Hz
- Range: 90 to 105 VAC, 50/60 Hz ±2 Hz
- Power Consumption: Ventilator 1.7 amp, ventilator and nebulizer 3.2 amp, ventilator, nebulizer and compressor 7.5 amp
Pneumatic Power
- Air and Oxygen: 30 to 90 psig (2.1 to 6.2 bars) @ 160 LPM
Physical Dimensions and Weight
(with computer and base enclosed)
- Dimensions: 54.5" height x 22.5" wide x 22.5" deep A(138 x 57 x 57 cm)
- Weight: 185 lbs (84 kg)