DRE Medical Equipment
englishespañol
phone: 1-800-462-8195DRE Facebook PageDRE LinkedIn PageDRE Twitter Profile

Refurbished - Ohmeda Care Plus Infant Incubator

Condition:
  • Refurbished
Market Availability:
Limited
download sales sheetSales Sheet
More Resources
Manuals, Docs, Links
Ohmeda Care Plus Infant Incubator

Product Description

Integrates microprocessor-based controls with a modular design for efficient patient care

  • Features a microprocessor that lets you control the patient environment in patient and air modes
  • Provides instant response for maintaining a precise thermoneutral environment, avoiding unsteady temperatures found on traditional analog systems

 

SAFETY FEATURES:

  • Multiple audio tones, including indicator LEDs for identifying alarm conditions
  • Alarms: Patient temperature, control temperature, high temperature, air circulation, probe failure, system failure, and power failure
  • Independent safety circuit



SPECIFICATIONS

Dimensions (incubator with cabinet)

  • Height: 53.5” (135.9 cm)
  • Depth: 25.5” (64.8 cm)
  • Width: 35” (88.9 cm)
  • Mattress: 13.7” x 25.6” (34.8 cm x 65 cm)
  • Tilting positions: ±8° in 4° angular increments
  • Weight: 185 lbs (84.04 kg)
  • Casters: 5” diameter, two locking, two nonlocking

Power Requirements

  • 120 VAC 60 Hz models: 115V ±10%; 5.7 amps
  • 220 VAC 50/60 Hz models: 220V ±10%; 3.0 amps
  • 240 VAC 50/60 Hz models: 240V ±10%; 2.7 amps
  • Nominal power consumption: 450 watts at maximum heater output
  • Line voltage compensation: Input voltage is monitored and heater output is adjusted to compensate for line voltage variation
  • Leakage current: Less than 40 microamperes on 120V units (18 microamperes on 220V and 240V units) measured at patient probe connection. Less than 90 microamperes on 120V units (180 microamperes on 220V and 240V units) measured at an exposed metal surface.

 

Performance Specifications

  • Temperature rise time: 25 to 30 minutes per 10° C
  • Temperature variability: 0.2° C
  • Temperature overshoot: 0.5° C
  • Air velocity over mattress: Less than 10 cm/sec (with inner wall option in place); 35 cm/sec (without inner wall option)
  • Noise level within unit: Less than 60 dB

 

Oxygen Concentration

  • 5 liters/minute oxygen input, range 25 to 35%
  • 10 liters/minute, 30 to 50%
  • 15 liters/minute, 40 to 70%

 

Humidification

  • Up to 50% ±10%

 

Patient Temperature Measurement System

  • Patient control (Servo) mode: 35° to 37° C (95° to 98.6° F)
  • Air control (Manual) mode: 20° to 37° C (73.4° to 98.6° F)
  • Up to 39° C (102.2° F) with control panel override switch
  • Resolution: ±0.1° C or F
  • Probe interchangeability: ±0.1° C (±0.2° F)