DRE Medical Equipment
englishespañol
phone: 1-800-462-8195DRE Facebook PageDRE LinkedIn PageDRE Twitter Profile
This item has been identified as hard to find, discontinued, or unavailable at this time. Although we will be happy to work with you in locating this item, please consider the alternatives listed below.

Baxter Colleague Volumetric Infusion Pump

Condition:
  • Retired
Baxter Colleague Infusion Pump

Product Description

Baxter's Colleague Volumetric Infusion Pump is designed to meet the fluid delivery needs of today's evolving healthcare environment.

The Colleague 3 Pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as, intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces applications.

Fluid delivery applications include:

  • Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
  • Whole blood and blood products.

The Colleague Volumetric Infusion Pump feature spectrum includes:

  • Micro and Macro Rate Range
  • Basic Delivery Programming
  • Piggybacking Secondary Medications
  • Special Programming functions for Dosing

Specifications

  • Ac power Requirements: 100-120 VAC, 60Hz
  • Leakage Current: Less than 300 u amps earth leakage (tested per UL 2601)
  • External Fuses: 1.6 amp Type T (time delay) 250V
  • Power Cord: 2.9 m (9 feet) long with integrally molded hospital grade plug
  • Battery Supply System: Colleague 3 Pump - 3 hours operating time at 100 mL/hr for all three channels (fully charged new batteries) 12 hours to recharge to 80% of capacity Internal charge system recharges batteries whenever the device is connected to an AC outlet Range of Programmable Flow Rates Primary Infusion on each pump channel 0.1 to 99.9 mL/hr in 0.1 mL/hr increments 1 to 1200 mL/hr in 1 mL/hr increments Piggyback Infusion on each pump channel 0.1. to 99.9 mL/hr in 0.1 mL/hr increments 1 to 500 mL/hr in 1 mL/hr increments Note: Rate limits can be configured for values less Than those stated above.
  • Volume to be Infused: 0.1 - 99.9 mL in 0.1 mL increments 1- 9999 mL mL/hr in 1 mL increments. Note: Volume to be Infused limits can be configured for values less than those stated above
  • KVO: 0.1 - 5 mL/hr in 0.1 mL increments (configurable option) or programmed rate whichever is less.
  • Priming rate: 500 mL/hr (to be used only when not connected to patient)
  • Advance Air Rate to Advanced an Air Bubbles: At the same rate programmed for the current primary or piggyback infusion.
  • Air Bubble Setting: Measured accumulations of approximately
  • Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
  • Whole blood and blood products.

Suggested Alternatives



TOP^

Recently Viewed Products




RSS