DRE offers the Respironics Novametrix Tidal Wave model 610 portable handheld pulse oximeter and CO2 monitor with SpO2. The Tidal Waves are the only handheld, mainstream capnographs available today, and provide immediate feedback on the patient's airway condition and ventilatory status. Based on the same technology as Respironics Novametrix bedside capnographs. THREE OPTIONAL CONFIGURATIONS: FEATURES: CAPNOGRAPH Initialization Time Calibration CAPNOSTAT® Mainstream CO2 Sensor CO2 (Carbon Dioxide) Respiratory Rate Sampling Flow Rate (Model 715 only) PULSE OXIMETER Principle of Operation SpO2 (Oxygen Saturation) Pulse Rate Sensors GENERAL Display Trending Capabilities Communication Output Physical Electrical Environmental
Handheld Mainstream Capnograph
Handheld Capnograph with Mainstream and Sidestream Sampling
Handheld Mainstream Capnograph/Pulse Oximeter Tidal Wave Sp
Handheld Mainstream Capnograph/Pulse Oximeter with Sidestream Sampling Capabilities
Principle of Operation
Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
Capnogram within 15 seconds, full specifications within 60 seconds.
No calibration required.
Size: 3.30 × 4.24 × 2.16 cm (1.30 × 1.67 × 0.85 in.).
Weight: 18 grams (0.64 oz.), cable excluded
Shock Resistance: Withstands repeated six foot drops to a hard surface.
Range: 0 - 150 mmHg
Accuracy:
±2 mmHg (for 0 - 40 mmHg),
±5% of reading (for 41 - 70 mmHg),
±8% of reading (for 71 - 150 mmHg)
Resolution: 1 mmHg
Response Time: 60 ms
Units: mmHg, kPa, or %
Compensations: N2O & O2 (selectable), Pb (automatic)
Capnogram: Selectable sweep speed
Accuracy: ±1 br/min
Range: 0-150 br/min
Resolution: 1 br/min
180 ml/min (for non-intubated patients)
Response time is less than 200 ms at 180 ml/min flow
Red/Infrared absorption.
Range: 0-100%
Accuracy: ±2% SpO2 (for 80-100% SpO2)
(1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
Resolution: 1%
Averaging: Menu selectable 2 or 8 seconds
Range: 30-250 bpm
Accuracy: ±1% of full scale
Resolution: 1 bpm
Averaging: 8 seconds
Reusable Y-Sensor (can be sterilized and used with all patient populations);
reusable adult finger sensor
Alerts
Limits:Adjustable limits for ETCO2, SpO2, Respiratory Rate and Pulse Rate. No Respiration Detected adjustable timeout.
Audio: Adjustable volume, two minute silence or off (LED indicators)
Visual: On-screen and LED
Type: Dot matrix, backlit (LCD)
Size: 2.2 × 1.1 inches
Memory: 24 hr., battery backed
Format: 30 minute on-screen or printed in trend or tabular mode.
Download data to personal computer for use with NovaCARD software
RS232 to Seiko DPU-414 printer, NovaCARD software or computer
Size (height, width, depth): 20 × 9.5 × 4.8 cm (8 × 3.75 × 1.875 in.).
Weight: 793 grams (28 oz.).
External Supply (optional)
Input: 120 VAC 60 Hz, or 220 VAC 50 Hz
Battery: Nickel-Metal-Hydride (NiMH): 4.5 hour life
Battery: Standard AA lithium: 5.5 hour life
Operating temperature: 0 to +40 °C (32-104 °F)
Operating humidity: 0 to 90% RH non-condensing