Add Condition:
Market Availability:
Limited
Product Description
FEATURES
- Monopolar and Bipolar Outputs
- Audible and Visual Arms, including REM(TM) Fault Indicator
- REM(TM) (Return Electrode Monitoring) (Standard)
- 300 Watts Output - Cut, 120 Watts - Coagulation, 70 watts - Biopolar
- Monopolar Foot Switch Included
- Bipolar Foot Switch (Optional)
- Rolling Carts (Optional)
SPECIFICATIONS
- Dimensions: 8 x 11 x 17 inches (15 x 28 x 3 cm)
- Weight: 15 lbs (6.4 kg)
Output Characteristics
- Cut - 2400 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 200 Nominal Power at rated load watts, 1.8 @ 100 W Crest Factor at rated load typical
- Blend 1 - 2800 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 175 Nominal Power at rated load watts, 2.7 @ 100 W Crest Factor at rated load typical
- Blend 2 - 3200 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 150 Nominal Power at rated load watts, 3.3 @ 100 W Crest Factor at rated load typical
- Blend 3 - 3600 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 125 Nominal Power at rated load watts, 4.6 @ 100 W Crest Factor at rated load typical
- Coag - 5000 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 75 Nominal Power at rated load watts, 8.0 @ 50 W Crest Factor at rated load typical
- Bipolar - 800 Max (open circuit) P-P Voltage, 100 Rated Load ohms, 50 Nominal Power at rated load watts, 1.8 @ 40 W Crest Factor at rated load typical
Output Waveform
- Cut - 500 kHz Sinusoid
- Blend 1 - 500 kHz bursts, 50% Duty Cycle, 31 kHz Recurrence
- Blend 2 - 500 kHz bursts, 37.5% Duty Cycle, 31 kHz Recurrence
- Blend 3 - 500 kHz bursts, 25% Duty Cycle, 31 kHz Recurrence
- Coag - 500 kHz bursts, 31 kHz Recurrence
- Bipolar - 500 kHz bursts, Unmodulated Recurrence
Low Frequency Leakage:
(50/60 Hz)
Normal polarity: intact chassis ground - <10 mA, ground open - <100 mA.
Reverse polarity: ground open - <100 mA. Sink current, 140V applied, all inputs - <150 mA. Chassis source current, ground open - <110 mA
High Frequency Risk Parameters
Bipolar and Monopolar RF leakage current: <150 milliamperes
REM™ Contact Quality Monitor:
Measurement frequency: 140 kHz ± 10 kHz.
Measurement current: 1.5 mA ± 0.5 mA
Input Power Source:
Nominal input line voltage is 120 VAC with a full range of 85-135 VAC. Frequency line varies from 45 to 65 Hz
- Idle - 0.5 A (Max. Current), 60 W (Max. Power)
- Cut - 5.0 A (Max. Current), 600 W (Max. Power)
- Coag - 30 A (Max. Current), 300 W (Max. Power)
- Bipolar - 1.5 A (Max. Current), 180 W (Max. Power)
TOP^Recently Viewed Products