Superior flexibility and performance with a premium flat panel display
The DRE Waveline Pro displays as many as eight colorful waveforms on a crisp 15” high resolution display. It is lightweight, portable and measures multiple parameters, including ECG (3 or 5 lead) with ST segment monitoring, resp., dual temp., SpO2 and NIBP. Configurations are available that measure capnography, dual channel IBP, and anesthetic agents, and that have a built-in thermal printer.
FEATURES
- New - Optional Plug-In Anesthesia Gas Module expands measurement capabilities -- Anesthesia Gas Module
- New - Bigger time stamp in upper right corner
- New - Standby button
- New - Revised SpO2 pitch tone is still audible when SpO2 rate decreases.
- New - Volume and Sounds are adjustable and now can be turned off completely
- New - Option to print Numerical Data Only
- New - Color of the waveforms can be changed
- Works with EMR and wireless telemetry systems
- Optional agent bench is compact and part of main unit. All in one monitor!
- Displays as many as eight waveforms
- Multi-lead simultaneous ECG monitoring
- Automatically set alarm limits
- Graphical and tabular trending
- Color-coded alarms
- Patient data entry
- Battery backup (two removable batteries)
- Networking capability
- Large (15”), high resolution flat panel display
- Available with a built-in thermal printer
SPECIFICATIONS
Physical Dimensions & Weight
- Base Unit: 12.1”/309 mm (H) x 14.4”/365 mm (W) x 6.3”/159 mm (D)
- Weight: 17.6 lbs/8.0 kgs (including anesthetic agents module and batteries)
Application
-
Neonatal, pediatric and adult patients
Performance Specifications
-
Display: 15” color TFT, Resolution 1024 x 768
-
Trace: 8 waveforms
-
Indicators:
Alarm indicator
Power indicator
QRS beep and alarm sound
-
Trend time: 1 – 48 hours
-
Recorder:
Built-in, thermal array, 2 channels
Record width: 1.9”/48 mm
Recorder paper: 2”/50 mm
Record speed: 25 mm/s, 50 mm/s
ECG
-
Input: 5-lead ECG cable and standard AAMI line for connection
-
Lead Choice: I, II, III, aVR, aVF, aVL, V, TEST
-
Gain Choice : x0.5, x1, x2, x4
-
Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
-
ECG Waveforms: 7 channels
-
Penetration Voltage: 4000VAC 50/60Hz
-
Sweep Speed: 12.5, 25 and 50 mm/s
-
HR Display Range: 30 – 300bpm
-
Accuracy: ±1 bpm or ±1%, whichever is greater
-
Alarm Limit Range Setting: Upper limit 100 – 200 bpm; Lower limit 30 – 100 bpm
RESP
-
Measure Method: RA-LL impedance
-
Range: 0 – 120 rpm
-
Accuracy: ±3 rpm
-
Alarm Limit Setting: Upper limit 6 – 120 rpm: Lower limit 3 – 120 rpm
-
Sweep Speed: 12.5 and 25 mm/s
TEMP
-
Range: 25° – 50° (C)
-
Accuracy:
± 0.2° C (25.0° – 34.9° C)
± 0.1° C (35.0° – 39.9° C)
± 0.2° C (40.0° – 44.9° C)
± 0.3° C (45.0° – 50.0° C)
-
Display Resolution: 0.1° C
-
Alarm Limit Setting: Upper limit 0° – 50° C, lower limit 0° – 50° C
-
Channel: 2 channels
SpO2
-
ASpO2: Anti-motion SpO2
-
SpO2% Range: 0 – 100%
-
SpO2 Accuracy: ±2% (70 – 100%, non-motion) ±3% (70 – 100%, motion)
-
Pulse Rate Range: 30 – 250 bpm
-
Pulse Rate Accuracy: ±2 bpm (non-motion) ±3 bpm (motion)
-
Alarm Limit Setting: Upper limit 70 – 100%, lower limit 70 – 100%
-
SpO2 Probe:
Red light LED wavelength – 660nm±5nm
Infrared light LED wavelength – 940nm±10nm
NIBP
-
Measuring Technology: Automatic oscillating measurement
-
Cuff Inflating: <30s (0 – 300 mmHg, standard adult cuff)
-
Measuring Period: AVE<40s
-
Mode: Manual, Auto
-
Measuring Interval in AUTO Mode: 2 min – 4 hrs
-
Pulse Rate Range: 30 – 250 (bpm)
-
Measuring Range:
Adult/Pediatric Mode
- SYS: 40 – 250 (mmHg) DIA :15 – 200 (mmHg)
Neonatal Mode
- SYS: 40 – 135 (mmHg) DIA : 15 – 100 (mmHg)
-
Resolution: 1mmHg
-
Accuracy:
Maximum Mean error: ±5mmHg
Maximum Standard deviation: 8mmHg
-
Overpressure Protection:
Adult Mode: 300(mmHg)
Neonatal Mode: 160 (mmHg)
-
Alarm Limit Setting:
- SYS: 50 – 240 mmHg
- DIA: 15 – 180 mmHg
IBP
-
Measurement Range: -50 – 300mmHg
-
Channel: 2 channels
-
Pressure Transducer: Sensitivity, 5μV/V/mmHg
-
Impedance Range: 300 – 3000Ω
-
Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
-
Unit: mmHg/kPa selectable
-
Resolution: 1 mmHg
-
Accuracy: ±1 mmHg or ±2%, whichever is greater
-
Alarm Range: -10 – 300 mmHg
EtCO2 (Dräger)
-
CO2 Measurement Range: 0 – 99 mmHg
-
Accuracy:
±2 mmHg (0 – 38 mmHg)
39-99 mmHg ±5% of reading + 0.08% for every 1 mmHg (above 38 mmHg)
-
Sampling Rate: 50 ml/min
-
Initialization Time: 30 seconds (typical), reaches ±5% steady-state accuracy within 3 minutes
-
Respiration Rate: 0 – 150 breaths/min
-
Mode: Adult, neonate
Anesthetic Agents
-
Method: Infrared absorption
-
Gas Sorts: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O (optional Automatic Agent ID)
-
Measurement Range:
Halothane, Isoflurane: 0 – 8.5%
Enflurane, Sevoflurane: 0 – 10%
Desflurane: 0 – 20%
CO2: 0 – 10%
N2O: 0 – 100%
-
Bias:
Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane: ±(0.15 Vol% + 15% rel.)
CO2: ±(0.5 Vol% + 12% rel.)
N2O: ± (2 Vol% + 8% rel.)
Networking
-
Industry standard 802.11b/g wireless network
Power
-
Source: External AC power or internal battery
-
AC Power: 100 – 240VAC, 50/60Hz, 150VA
-
Battery: Built-in and rechargeable
-
Charge Time: 4 hours
Environmental Specifications
-
Temperature:
Operating: 5° – 40° C
Storage: -20° – 65° C
-
Humidity range:
Operating: ≤80 %
Storage: ≤80 %
| Part Name and Description | DRE Part # |
| Blood Pressure Cuffs | |
| Blood Pressure Cuff Hose | INF-20020 |
| DRE Cloth 2-Piece Blood Pressure Cuffs w/Integrated Bladder - Single Lumen, Female Luer Connector | multiple |
| Trimline Disposable Neonatal Blood Pressure Cuffs - Single Lumen, Male Luer Slip Connector | multiple |
| CO2 / Agent Accessories | |
| CO2 Adult Airway Adapter Kit with Dehumidifier (Pack of 10) | INF-5004AAK4 |
| CO2 Sampling Line (Pack of 10) | INF-5001SL |
| External CO2 Module | INF-5CO2MR |
| Mainstream Disposable Adult Airway Adapter (Box of 10) | INF-5006SPAM |
| Mainstream Disposable Neonatal Airway Adapter (Box of 10) | INF-5006SNAA |
| Mainstream Reusable Adult Airway Adapter | INF-5006RPAA |
| Mainstream Reusable Neonatal Airway Adapter | INF-5006RNAA |
| Moisture Trap for Anesthetic Agent (5 each) | INF-5007MTD |
| Sample Line for Anesthetic Agent | INF-5006SC |
| Sample Line with Cannula for Anesthetic Agent (Pack of 5) | INF-5007SCA |
| Straight T Adapter | INF-5005TA |
| ECG Cables | |
| 3-Lead ECG Cable and Leads | INF-1003L6P0 |
| 5-Lead ECG Cable and Leads | INF-1005L6P0 |
| ECG Electrodes | |
| Adult Snap Electrodes Disposable (1,200/Box) | INF-100EA |
| Disposable Pediatric Snap Electrodes (1,500/Box) | INF-100EP |
| IBP | |
| Disposable IV Transducer (Pack of 5) | INF-6119 |
| IBP Cable | INF-6001IPC |
| Patient Monitor Batteries | |
| Battery for DRE Waveline EZ | INF-9200 |
| Battery for DRE Waveline PRO & Waveline Touch | INF-9100 |
| Power Cord | |
| 110V Power Cord | INF-9001 |
| 220V Power Cord | INF-9002 |
| Printer & Printer Paper | |
| Printer for Waveline Monitors | INF-700P03 |
| Printer Paper for Waveline Pro, EZ, Touch and Plus (5 Rolls) | INF-700PP0.5 |
| Rolling Stands & Wall Mounts | |
| Base Plate for Waveline EZ | INF-8004BP |
| Base Plate for Waveline PRO | INF-8003BP |
| Base Plate for Waveline Touch | INF-8002BP |
| Rolling Stand with Basket | INF-800RS |
| Wall Mount | INF-800WM |
| SpO2 Accessories | |
| Digital Adult Oximetry Sensor With Direct Connect | INF-325DFS |
| Digital Adult Oximety Sensor for Extension Cable | INF-325DFSEXT |
| Digital pediatric Oximetry Clip for Direct Connect | INF-345DFSPC |
| Digital Pediatric Oximetry Clip for Extension Cable | INF-340DFSPC |
| Digital Y sensor for Extension Cable | INF-325DYS |
| Disposable Digital Adult Oximetry Sensor for Extension Cable | INF-330DDFSA |
| Disposable Digital Neonatal Oximety Sensor for Extension Cable | INF-3300DDFSN |
| Disposable Digital Pediatric Oximety Sensor for Extension Cable | INF-3300DDFSP |
| SpO2 Extension Cables | |
| Digital Oximetry Extension Cable | INF-325DEC |
| Temperature Probes | |
| Oral / Rectal Temperature Probe | INF-400T0R |
| Skin Surface Temperature Probe | INF-400TS |