Datex-Ohmeda 5330's are suitable for performing anesthetic vaporizer concentrations checks in the field. These units are available either used or professionally refurbished from DRE.
Features
• Monitors 5 Agents: Enflurane, Halothane, Isoflurane, Sevoflurane, Desflurane
• Easy-to-read digital display screen 5330
Specifications
Dimensions / Power
Height: 4 in, 10.16 cm
Width: 10 in, 25.4 cm
Depth: 16 in, 40.64 cm
Weight: 13.25 lbs, 6 kg
Power Input: 110, 120 VAC, 50/60 Hz
Operational Environment
Temperature: 15 to 35ºC
Humidity: 0 to 95% RH, non-condensing
Storage Environment
Temperature: -5 to 50ºC
Humidity: 0 to 95% RH, non-condensing
Display
Displayed Parameter: Agent inspired & expired concentration levels (%)
Trendgraph: Last 120 breaths, 20 min, 60 min, 120 min
Sample Flow Rate: 250 mL/min
Response Time: 600 msec
Warm-Up Time: 5 minutes to operation
Compensation: Temperature & barometric pressure
Analog Output: Anesthetic agent waveform
Digital Output: RS-232 compatible
Enflurane, Halothane, Isoflurane, Sevoflurane
Range: 0 to 9%
Accuracy: 0 to 1.0% (+/- 0.2%)
1.1 to 6.0% (+/- 0.3%)
6.1 to 7.0% (+/- 0.35%)
Resolution: 0.1%
Alarms: High expired 0.2 to 10% agent concentration
High inspired 0.2 to 10% agent concentration
Low expired 0 to 9.8% agent concentration
Low inspired 0 to 9.8% agent concentration
Desflurane
Range: 0 to 24%
Accuracy: 0 to 4.9% (+/- 0.2%)
5.0 to 9.9% (+/- 0.5%)
10.0 to 18.0% (+/- 1.0%)
18.1 to 22.0% (+/- 2.0%)
Resolution: 0.1%
Alarms: High expired 0.2 to 10% agent concentration
High inspired 0.2 to 10% agent concentration
Low expired 0 to 9.8% agent concentration
Low inspired 0 to 9.8% agent concentration