FEATURES
Warranty
TECHNICAL SPECIFICATIONS
Power requirements: Obtained from rechargeable 120V battery
Power required to recharge battery: 110 to 125 VAC, 50/60 Hz, 6.0 amps
Battery capacity: Approximately 10,000 mAs at 100 kVp, when fully charged
Motor drive ratings: The drive assist unit can be driven at speeds up to 180 feet per minute or 54.9 meters per minute
X-ray generator ratings and duty cycle: All radiographic exposures with this unit are at 100 milliamperes
| kVp (maximum) | mAs (maximum) |
| 110 to 125 | 200 |
| 95 to 105 | 250 |
| 50 to 90 | 320 |
All rates and duty cycles are subject to limitations of the X-ray tube rating and apply at altitudes up to 8,000 feet (2,438 kilometers), average relative humidity not exceeding 80% and ambient temperature between 60º F (15.6º C) and 100º F (37.8º C)
Typical accuracies: ±6% of kVp and ±10% of mAs
| Selection and Range | Accuracy |
| 0.64 mAs to 1.25 mAs | ±30% |
| 1.6 mAs to 10 mAs | ±20% |
| 12.5 mAs to 320 mAs | ±12% |
| 50 kVp to 74kVp | ±(9% + 1kV) |
| 76 kVp to 125 kVp | ± (12% + 2 kV) |
kVp accuracies subject to additional ±(5% + 2kV) and mAs to ±3% accuracy of instrumentation used for calibration and measurement.
Conditions to meet previously mentioned accuracies
Collimator Specifications