Refurbished - Critikon 8720 Multi-Parameter Monitor

Add
Condition:
Refurbished
Market Availability:
Limited
Send to a friend
Critikon Dinamap 8720

Product Description

Features:

Now the DINAMAP® Compact Monitor is available in four different configurations offering you the technology that's right for you.

Each model features our unique DINAMAP® Technology, the "Gold Standard" in BP accuracy. Optional, integrated printers are also available in each model.

  • Suitable for Adult, Pediatric and Neonatal Use
  • Maintains Accuracy even when the Patient Moves
  • Allows you to take all Connected Vitals with the Press of One Button
  • STAT Mode for Rapid BP Readings in Emergency Situations
  • Single Determinations or Automatic Monitoring of Blood Pressure
  • User-Adjustable Target Inflation Pressure
  • Oral / Rectal Predictive Temperature
  • Built-In Printer Available

Specifications: 

Noninvasive Blood Pressure

  • Determination Time:  Automatic and Manual Modes, 20 to 45 seconds typical (120 seconds max); STAT mode, 19 seconds typical at nominal blood pressure and heart rate
  • Range:  Systolic, 30-245 mmHg; Diastolic, 10-210 mmHg; MAP, 20-225 mm Hg; Heart rate, 40-200 bpm
  • Alarm Limits:  systolic - high, 75-245 mmHg; low, 30-150 mmHg; diastolic - high, 50-210 mmHg; low 10-120 mmHg; MAP - high, 70.225 mmHg; low, 20-120 mmHg
ECG
  • Heart Rate Range:  10-250 bpm
  • Connector:  AAMI recommended, ECGC; MS3102A-14S-6S
  • Leads:  Selectable, I, II, III Lead off detection
  • Band Width:  0.5-40 Hz
  • Sweep Speed:  12.5, 25 and 50 mm per second
  • Gain:  1mV to 8mV full scale, selectable in 5 steps
  • QRS Detection Tone:  Volume selectable from 0(off) to 15
  • Alarm Limits:  Heart rate-high, 80-250 bpm; low, 10-140 bpm
Oxygen Saturation:
  • Accuracy:  (+/- standard deviation) ±2% at 90-99%; ±3% at 75-89%; unspecified under 75%
  • Saturation Range:  40-99%
  • Pulse Rate:  From 30-250 bpm
  • Pulse Averaging:  User selectable for 0 to 30 sec. in increments of 3.  Preset at 6 seconds at power turn on.
  • Saturation Pitch Indicator:  Pitch changes with saturation.  Volume selectable from 0(off) to 15.
  • Alarm Limits:  Saturation - high, 90-99%; low, 40-95%
Alarm Functions:
  • System:  Internal system test performed at power-up and during operation
  • Volume:  Adjustable from 1 to 15 on all parameters
  • NIBP:  Procedural alarm for overpressure, excess determination time, excess pump up time, no determination
  • SpO2:  Procedural alarms for low signal, probe disconnect, probe failure
 Mechanical:
  • Weight:  9.8 lbs
  • Dimensions:  9.25"H x 720"W x 8.5D
  • Display:  Electroluminescent, 320 x 256 pixels; viewing area, 3.8"W x 3.0"H
  • Power Requirements:  88-268 VAC; 47-63 Hz
  • Fuse Requirements:  2.0 amps
  • Leakage Current: <100µA

 




RSS