The 3800 Pulse Oximeter has an easy-to-use interface and intuitive feature set, as well as advanced TruTrak® + technology, a proprietary clinical motion algorithm.
Features:
Physical Specifications:
Accuracy Specifications:
SpO2 -
Pulse Rate -
- Calibration: Functional
- Range: 0 to 100%
- Accuracy (1 standard deviation): 70 to 100% (±2 digits); 70 to 100% (±3 digits during conditions of clinical patient motion with TruTrak+ enabled); Below 70% (unspecified)
- Resolution: 1%
Relative Perfusion Index (PIr) pulsatile value -
- Range: 30 to 255 beats per minute (bpm)
- Accuracy (assuming a constant pulse rate): ±2% or ±2 bpm (whichever is greater)
- Accuracy (during conditions of clinical patient motion): Unspecified
- Resolution: 1 bpm
- Range: 0.00 to 9.99
- Averaging interval: 12 seconds
- Resolution: 0.01
Operational Specifications:
Alarm Limits:
- Low SpO2 = OFF or 50 to 100& (preset 85%)
- High SpO2 = 50 to 100% or OFF (preset OFF)
- Low pulse = OFF or 30 to 235 bpm (preset: adult = 40 bpm, neonate = 100 bpm)
- High pulse = 30 to 235 bpm or OFF (preset: adult = 130 bpm, neonate = 200 bpm)
Alarm Silence:
- 2 minute silence (automatic, 2 min. on start up), all mute capability
Signal Validation:
- Relative Perfusion Index pulsatile value PerfTrak® waveform display
Memory:
- 12 hour data storage for SpO2, pulse rate and alarm conditions (1 data point every 6 seconds)
SpO2 Calculation:
Display:
- 30 values per second/60 Hz
- 25 values per second/50 Hz
- Digital - LED
- Waveform - Backlit LCD
SpO2 Averaging Modes:
- Selectable 3, 6 or 12 seconds (short, medium, long) preset = 12 seconds (long)
Diagnostics:
- Automatic self test on start up, continuous memory tests during operation
Outputs:
- Serial (SpO2, pulse rate, alarms, error messages and trend data)
Electrical Specifications:
- Battery: 8 volt, 3.2 ampere-hours sealed lead-acid. Continuous charge while operating on AC
- Charge Time: 100% capacity ≈ 8 hours; 80% capacity ≈ 4 hours
- Operation Time: ≈ 5.5 hours on fully charged battery
- Low Battery Indicator: "Low Battery" displayed when < 5 minutes remaining
- Overload Protection: 2 ampere 250V fuses
- Power Consumption: 15 watts typical
- Input Voltage: 100 to 240 V, 50/60 Hz