Caring for low birth weight neonates requires creating the best possible thermal-neutral environment. The C2000 Infant Incubator helps to promote and foster neonatal care by delivering a heated and humidified environment.
Features:
Technical Specifications:
Air-Shields Incubator with VHA Stand -
- Width: 38.0" (99 cm)
- Depth: 26.3" (67 cm)
- Height: 52.5" - 60" (133.4 to 152.4 cm)
- Weight: 198 lbs (90 kg)
Air-Shields Incubator with Fixed Height Stand -
- Width: 46.8" (119 cm)
- Depth: 26.3" (67 cm)
- Height: 56" (142 cm)
- Weight: 175 lbs (79.5 kg)
Hood Specifications:
- Mattress tray size: 16" x 32" (40.6 81 cm)
- Mattress to hood: 16.25" (41.2 cm)
- Access panel opening: 11" (28 cm)
- Standard hood includes: 4 port doors; 2 iris ports; front access panel; 3 left and 3 right tubing grommets-front
- Mattress tilt: ±12° (±1° C)
- Micro air intake filter: 99.9% efficiency
- Particle size removal: 0.3 micron
- Air mode control range: 20 to 39.0°C
- Air mode override: >37.0°C
- Air flow velocity across mattress: <10 cm/sec
- Skin mode control range: 34.0 to 38.0° C
- Skin mode override: >37.0° C
- Temperature rise time (at 22°C ambient): <35 minutes
- CO2 levels: <0.5% when delivered at 750 ml/min at a point 10 cm above the mattress
- Operating noise level in hood: 47 dBA (49 dBA with Servo Controlled Oxygen)
- Casters (4 casters, 1 steering): 5" (12.7 cm)
- RS-232 output
- Controller with LCD with brightness control
- Selectable color combinations: white on blue background (default) or yellow on black background
- Dual-skin temperature monitoring
- 6 AC (Mains) outlet
- Keypad lock
- 24 Hour trend of: Air temperature, skin temperature (1 and 2), Relative humidity, Oxygen concentration, Heater power
Environmental Requirements -
- Operating temperature range: 20°C to 30°C ambient
- Storage temperature range: -25°C to 60°C ambient
- Operating humidity range: 5% RH to 99% RH, non condensing
- Storage humidity range: 0% to 99% RH, non condensing
- Operating RH sensor temp. range: 20°C to 41°C
- Operating O2 sensor temp. range: 20°C to 41°C
- O2 sensor calibration pressure: 600 to 900 mm Hg
Alarms -
High/Low Skin Temperature Alarms:
- Temperature (limiting ≤): 38°C for air temp. < 37°C; 40.0° C for air temp. > 37° C
- Baby set temperature: ±1° C (or ±0.5° C)
- Air set temperature: +1.5° C/-2.5° C
- High/low air temperature;
- High temperature cut-out;
- Air probe failure;
- Skin probe failure/disconnect;
- Low air flow;
- Low humidity (humidity option);
- Low/high oxygen (±3%) (oxygen option);
- Oxygen calibration required (oxygen option);
- Oxygen cell difference (oxygen option);
- Oxygen solenoid failure (oxygen option);
- Scale disconnect (scale option);
- Too much weight (scale option)
System Alarms:
Servo Humidity Option:
- Controller failure
- Sensor module failure
- sensor module out of position
- Motor failure
- Power failure
- Heater failure
- Sensor disconnected
- Key stuck
- Water capacity: 1000ml
- Refill time: >24 hrs @ 85 RH%;
- Range: 30 to 95%
- Display accuracy: ±6% RH%
Servo Oxygen Option:
- Accuracy: 100% cal ±3%
- Accuracy: 21% cal ±5%
- Range: 21 to 65%
- Resolution: 1%
Scale Option:
- Weight range: 0 to 15.4 lbs (0 to 7 kg)
- Weight display resolution: 1 g or 1 oz (OIML = 10 g or 1 oz)
- Weight accuracy: 2 g ± 1/2 digit up to 2 kg (OIML = 10 g); 5 g ± 1/2 digit over 2 kg
- Weight gain or loss trended over 7 days
Other Options (Attach to Accessory Rail):
- IV Pole weight limit: 10.4 lb (4.72 kg)
- Monitor shelf weight limit: 25 lb (11.4 kg)
- Rear access door: Provided with 2 port doors and 4 additional hood grommets 2 left and 2 right
- Ventilator tube support
- Oxygen tank mount
Drawer Options:
- Large deep swivel type
- Two shallow swivel out type