Refurbished - Air-Shields C-2000 Infant Incubator

Condition:
  • Refurbished
Market Availability:
Limited
download sales sheetSales Sheet
More Resources
Manuals, Docs, Links
Refurbished - Air-Shields C-2000 Infant Incubator

Product Description

Caring for low birth weight neonates requires creating the best possible thermal-neutral environment. The C2000 Infant Incubator helps to promote and foster neonatal care by delivering a heated and humidified environment.

Features: 

  • Optional Dual Access Doors – Allow access from either side, expediting care. 
  • Thermal Performance – Patented dual air curtains reduce radiant heat loss from the infant by warming the inner hood surface. A bi-directional airflow feature helps to minimize temperature fluctuations when the access doors are opened.
  • Large Mattress and Dual Probes – The C2000 incubator allows for simultaneous care of twins in the same unit, allowing more incubators for more patients.
  • Internal Noise Level less than or equal to 47dBA – Maintains a supportive environment for both patient and caregiver.
  • Optional In-bed Scale – Weigh the baby in a warm environment with minimal disturbance.
  • 24-hour Data Trends – This unit facilitates record keeping for air and skin temperature, oxygen, humidity, heater power, and weight changes.

Technical Specifications:

Air-Shields Incubator with VHA Stand - 

  • Width:  38.0" (99 cm)
  • Depth:  26.3" (67 cm)
  • Height:  52.5" - 60" (133.4 to 152.4 cm)
  • Weight:  198 lbs (90 kg)
Air-Shields Incubator with Fixed Height Stand - 
  • Width:  46.8" (119 cm)
  • Depth:  26.3" (67 cm)
  • Height:  56" (142 cm)
  • Weight:  175 lbs (79.5 kg)
Hood Specifications:
  • Mattress tray size:  16" x 32" (40.6  81 cm)
  • Mattress to hood:  16.25" (41.2 cm)
  • Access panel opening:  11" (28 cm)
  • Standard hood includes:  4 port doors; 2 iris ports; front access panel; 3 left and 3 right tubing grommets-front
  • Mattress tilt:  ±12° (±1° C)
  • Micro air intake filter:  99.9% efficiency
  • Particle size removal:  0.3 micron
  • Air mode control range:  20 to 39.0°C
  • Air mode override: >37.0°C
  • Air flow velocity across mattress:  <10 cm/sec
  • Skin mode control range:  34.0 to 38.0° C
  • Skin mode override:  >37.0° C
  • Temperature rise time (at 22°C ambient):  <35 minutes
  • CO2 levels:  <0.5% when delivered at 750 ml/min at a point 10 cm above the mattress
  • Operating noise level in hood:  47 dBA (49 dBA with Servo Controlled Oxygen)
  • Casters (4 casters, 1 steering):  5" (12.7 cm)
  • RS-232 output
  • Controller with LCD with brightness control
  • Selectable color combinations: white on blue background (default) or yellow on black background
  • Dual-skin temperature monitoring
  • 6 AC (Mains) outlet
  • Keypad lock
  • 24 Hour trend of: Air temperature, skin temperature (1 and 2), Relative humidity, Oxygen concentration, Heater power
Environmental Requirements -
  • Operating temperature range:  20°C to 30°C ambient
  • Storage temperature range:  -25°C to 60°C ambient
  • Operating humidity range: 5% RH to 99% RH, non condensing
  • Storage humidity range: 0% to 99% RH, non condensing
  • Operating RH sensor temp. range: 20°C to 41°C
  • Operating O2 sensor temp. range:  20°C to 41°C
  • O2 sensor calibration pressure:  600 to 900 mm Hg
Alarms -
  • Temperature (limiting ≤): 38°C for air temp. < 37°C; 40.0° C for air temp. > 37° C
  • Baby set temperature:  ±1° C (or ±0.5° C)
  • Air set temperature:  +1.5° C/-2.5° C
High/Low Skin Temperature Alarms: 
  • High/low air temperature;
  • High temperature cut-out;
  • Air probe failure;
  • Skin probe failure/disconnect;
  • Low air flow;
  • Low humidity (humidity option);
  • Low/high oxygen (±3%) (oxygen option);
  • Oxygen calibration required (oxygen option);
  • Oxygen cell difference (oxygen option);
  • Oxygen solenoid failure (oxygen option);
  • Scale disconnect (scale option);
  • Too much weight (scale option)
System Alarms:
  • Controller failure
  • Sensor module failure
  • sensor module out of position
  • Motor failure
  • Power failure
  • Heater failure
  • Sensor disconnected
  • Key stuck
Servo Humidity Option: 
  • Water capacity:  1000ml
  • Refill time:  >24 hrs @ 85 RH%;
  • Range:  30 to 95%
  • Display accuracy: ±6% RH%
Servo Oxygen Option:  
  • Accuracy:  100% cal ±3%
  • Accuracy:   21% cal ±5%
  • Range:  21 to 65%
  • Resolution: 1%
Scale Option:
  • Weight range:  0 to 15.4 lbs (0 to 7 kg)
  • Weight display resolution:  1 g or 1 oz (OIML = 10 g or 1 oz)
  • Weight accuracy:  2 g ±  1/2 digit up to 2 kg (OIML = 10 g); 5 g ± 1/2 digit over 2 kg
  • Weight gain or loss trended over 7 days
Other Options (Attach to Accessory Rail):
  • IV Pole weight limit:  10.4 lb (4.72 kg)
  • Monitor shelf weight limit: 25 lb (11.4 kg)
  • Rear access door:  Provided with 2 port doors and 4 additional hood grommets 2 left and 2 right
  • Ventilator tube support
  • Oxygen tank mount
Drawer Options:
  • Large deep swivel type
  • Two shallow swivel out type 

 

TOP^