Product Description
The BEAR CUB 750vs Infant Ventilator raises the standard
of care with unparalleled value. and is available professionally
refurbished by DRE. -- En Espanol
The Bear Cub 750vs infant ventilator is designed to
accommodate a range of patients from the small neonate to the 30kg
pediatric. The Bear Cub 750vs and 750psv Infant Ventilators are designed to
provide ventilator support for the critical care management of patients ranging
from the small neonate (500 grams or larger) to the small pediatric patient (up
to 30 kg) with compromised lung function. The 750vs has integrated
synchronized ventilation and tidal volume monitoring. It also includes
independent base & inspiratory flows and an internal battery
back-up.
The Bear Cub 750 Infant Ventilators
offer:
- Integrated Synchronized Ventilation
- Tidal Volume Monitoring
- Independent Base & Inspiratory Flows
- Internal Battery Back-Up
- Exclusive Volume Limit* Feature
- Assist Control
- Flow Cycled AC
- SIMV/IMV
- Flow Cycled SIMV
- CPAP
- SIMV with Pressure Support
- Pressure Support Ventilation
Specifications:
Controls
- Mode: AC, Flow cycle AC, SIMV/IMV, Flow Cycled SIMV,
SIMV/PSV, PSV, CPAP
- Ventilator Rate: 1.0 to 150 BPM
- Inspiratory Time: 0.1 to 3.0 sec.
- Inspiratory Flow: 1.0 to 30 LPM
- Base Flow: 1.0 to 30 LPM
- Volume Limit: 5.0 to 300 ml
- PEEP: 0 to 30 cmH2O
- Inspiratory Pressure: 0 to 72 cmH2O
- Manual Breath: x One
- Assist Sensitivity: 0.2 to 5.0 LPM; Minimum to Maximum
bar
- Over Pressure Relief: 15 to 75 cmH2O
- % O2: 21 to 100%
- Alarm Loudness: Min 60 to Max. 75 dbA
- Audible Off Volume Limit: LED On/Off
Alarms
- High Breath Rate: 3.0 to 255 BPM
- Low PEEP/CPAP: -5.0 to 30 cmH2O
- Low Inspiratory Pressure (vs): 1.0 to 65
cmH2O
- Low Inspiratory Pressure (psv): 5 to 41 cmH2O (derived
from High Pressure Limit and Low PEEP/CPAP Alarm Settings
- Patient Circuit: LED On/Off
- Failed to Cycle: LED On/Off
- Low Gas Supply: LED On/Off
- Apnea Alarm: 5, 10, 20, 30 sec
- Settings Incompatible: LED On/Off
- Pressure Settings Incompatible: LED On/Off
- Prolonged Inspiratory Pressure: LED On/Off
- Flow Sensor: LED On/Off
- Low Battery: LED On/Off
- Alarm Silence: 60 sec
- Visual Reset: Push Button
- High Pressure Limit: 10 to 75 cmH2O
- Low Minute Volume: Off to 99 lpm
Monitors
- PEEP: 0 to 30 cmH2O
- Breath Rate: 0 to 255 BPM
- Breath Type (Patient Initiated): LED
- Minute Volume: 0 to 30.0 LPM
- Tidal Volume (Exhaled): 0 to 500 ml
- % Tube Leak: 0 to 100%
- Inspiratory Time: 0.1 to 3.10 sec Expiratory Time 9 to 99.9 sec
- I:E Ratio: 9.9:1 to 1:9.9
- Peak Inspiratory Pressure: 0 to 99 cmH2O
- Mean Airway Pressure: 0 to 75 cmH2O
- Air Pressure: 0 to 100 psig; 0 to 99 psig
- O2 Pressure: 0 to 100 psig; 0 to 99 psig
- Proximal Airway Pressure: -10 to 100 cmH2O
- Hour meter: 0 to 99,999 hours
- Test: Push Button
- Line Power: Green/Red LED
- Battery: LED On/Off
- Inspired Tidal Volume: 0 to 500 ml
Limits to Ventilation (Fixed)
- Minimum Expiratory Time: 0.15 seconds
- Maximum Inspiratory Time: 3:10 seconds
- Maximum settable I:E Ratio: 4:1
Tolerances:
Monitors
- Breath Rate: ± 1 BPM or ± 20 msec, whichever is
greater
- Inspiratory Time: ± 0.02 sec
- Expiratory Time: ± 0.02 sec
- I:E Ratio: ± 0.1 or ± 20 msec (on the calculation based on the monitored
inspiratory and expiratory times), whichever is greater
- Airway Pressure (Continuous):
± 1 cmH2O (-10 to 20 cmH2O)
± 2 cmH2O
(20 to 65 cmH2O)
± 3 cmH2O (65 to 100 cmH2O)
- Peak Inspiratory Pressure: ± 2 cmH2O or ± 5%, whichever is greater
- Mean Airway Pressure: ± 2 cmH2O or ± 3%, whichever is greater
- PEEP: ± 2 cmH2O or ± 5%, whichever is greater
- Air/O2 Pressure: ± 5 psig
- Minute Volume Accuracy: Based on the Tidal Volume and Breath Rate Monitors
- Inspiratory Tidal Volume: ± 1 ml or ± 10%, whichever is greater
- Expiratory Tidal Volume: ± 1 ml or ± 10%, whichever is greater
- % Leak: ± 1 ml or ± 2%, whichever is greater
(when compared to the
Inspired and Expired Tidal Volume Monitors)
- Hour Meter: ± 2% of reading
Alarms/Limits
- Low PEEP/CPAP: ± 2 cmH2O
- High Breath Rate: ± 1 BPM
- Low Minute Volume:
± 10 ml/min (0 to 99
ml/min)
± 0.1 L/min (0.1 to 9.9 L/min)
- High Pressure Limit: ± 4 cmH2O
- Prolonged Inspiratory Pressure: ± 2 cmH2O
- Volume Limit: ± 2 ml or ± 10% of setting,
whichever is greater
- Apnea: ± 1 sec
- Maximum (Pop-off) Pressure: ± 4
cmH2O
Controls
- Ventilator Rate: ± 1 BPM or ± 20 msec (applied
to the breath interval), whichever is greater
- Inspiratory Rate:
± 0.020 sec (0.10 to 0.50
sec)
± 0.025 sec (0.50 to 3.00 sec)
- Assist Sensitivity: Not a calibrated scale
- PEEP/CPAP and Inspiratory: ± 1 cmH2O or ± 5%,
whichever is greater (breath to breath)
- Base/Inspiratory Flow: ± 0.5 LPM or
± 10% of setting, whichever is greater (0 to 60 cmH2O)
- Oxygen %: ± 3%
- Degree of Safety in Presence of Flammable Anesthetics: None
- Mode of Operation: Continuous
Breathing Circuit Parameters
- Inspiratory Resistance: 0.42 cmH2o/l/min
- Expiratory Resistance: 0.08 cmH2o/l/min
- Compliance: 1.34 ml/cmH2o
- Internal Volume: 1370 ml
Outputs
- Digital: RS-232 Bi-Directional
- Analog: See Manual
- Proximal Pressure: -10 to 100 cmH2O, 1 cmH2)/25 mV
- Proximal Inspiratory/Expiratory Flow: -40 (expiratory) to 40 (inspiratory)
LPM, 1 LPM/50 mV
- Machine Delivered Flow: 0 to 30 lpm, 1 LPM/50 mV
- Breath Phase: 0.5V Logic Signal
- Remote Nurse Call: 0.5 amps max at 24 vdc
- Auxiliary Blended Gas Outlet: 7 to 17 psig, 0 to 8 LPM
Inputs
Electrical
- Voltage:
100 V - 80 to 110 VAC
120V - 96 to 132
VAC
220V - 176 to 242 VAC
240V - 192 to 264 VAC
- Current:
100V - 1.0 A maximum
120V - 1.0 A maximum
220V - 0.5 A
maximum
240V - 0.5 A maximum
- Frequency:
100V - 50.60 Hz
120V - 50/60 Hz
220V - 50/60 Hz
240V
- 50/60 Hz
- Fuses:
100/120V - T o.5 A, 5 x 20 mm
230/240V - T 0.5 A, 5 x 20
mm
(For 220V operation, set the power entry module seclector switch to 230V
setting)
Pneumatic:
- Oxygen and Air:
30 to 80 psig, 50 LPM Supply
Ratings
KPa (206 to 551)
Physical Dimension and Shipping Information:
- Ventilator Weight: 12 kg / 13.6 kg
- Graphics Display Weight: 2.5 kg
- Pedestal Stand Weight: 4.1 kg
- Compressor Weight: 50 kg
- Ventilator Dimensions: 13.5"W x 10"D x 11"H
- Graphics Display Dimensions: 13"W x 2.5"D x
9.5"H
- Pedestal Stand Dimensions: 24" x 40.5"H
- Compressor Dimensions: 22"W x 21.5"D x
36"H
- Shipping Weight Including Ventilator: 17 kg; 18.2
kg
- Graphics Display: 5 kg
- Pedestal Stand: 5.1 kg
- Compressor: 9.1 kg
- Shipping Dimensions
Ventilator: 18"W x 18"D x
19"H
Graphics Display: 16"W x 13"D x 11"H
Pedestal Stand: 25"W x 5"D x
45"H
Compressor: 26"W x 22"D x 40"H
Environmental Specifications:
Temperature
- Storage and Shipping: -40 to 70 C
- Checkout and Operating: 10 to 40 C
Altitude
- Checkout, Operating, Transport and Storage 0 to 3,000
meters (14.7 to 10.5 PSIA/760 to 543 mm Hg)
Humidity
- Storage and Shipping: 0 to 99% Relative Humidity -
Non-condensing
- Checkout and Operating: 0 to 95% Relative Humidity -
Non-condensing
Complete Specifications are listed in PDF format under the documents
tab.
TOP^Helpful Resources