Refurbished - Hewlett Packard Codemaster XL Defibrillator

Model Number:
71723
Condition:
  • Refurbished
download sales sheetSales Sheet
More Resources
Manuals, Docs, Links
Refurbished - Hewlett Packard Codemaster XL Defibrillator
Refurbished - Hewlett Packard Codemaster XL Defibrillator

Product Description

Hewlett Packard Codemaster XL and XL+ defibrillators are available professionally refurbished from DRE with a built-in recorder, ECG, and AC power with battery back-up.

OPTIONAL

  • SPO2 - Pulse Oximetry
  • Pacing
  • Advisory Mode
  • External electrode adapters.

SPECIFICATIONS

  • Dimensions
    7.9"H x 11.8"W x 15.6"L
    20cm x 30cm x 39.7cm

  • Weight:   24 lbs. / 10.9kg

  • Battery
    Type:   Rechargeable, sealed lead-acid battery.  4 Ah, 12 V nominal.
    Charge Time:  2 hours to 90% of full capacity.  18 hours to 100% capacity.  Repeated charging to less than 100% will reduce the useful life of the battery.
    Capacity:  2.5 hours monitoring or 50 full-energy discharges or 1 hour monitoring and recording.
    AC input:   100 to 230 VAC +/- 15%, 50/60 Hz. 
    Battery Indicators:  Illuminated LED indicates battery is charging. Low Battery message appears on the monitor.

  • Defibrillator
    Waveform:  damped sinusoidal.
    Output energy (delivered): 2,3,5,7,10,20,30,50,70,100,150,200,300, and 360 joules.
    Charge control:  push-button on apex paddle and on front panel.
    Charge time:  less than 5 seconds to 360 joules with a functional battery installed.
    Charge done indicator:  display indicates available energy, charge-done light on apex paddle, charge done tone.
    Synchronizer:  SYNC message appears on the monitor and is annoted periodically on the recorder while in synchronous mode.  An audible beep sounds with each detected R-wave, while a marker on the monitor and sync mark on the recorder strip indicate the discharge point.
    Paddles: standard paddles are anterior / anterior.  Adult electrodes (83 cm sq) slide off to expose pediatric electrodes (21 cm sq). Paddle cord is 10 feet (3m).  Full range of internal paddles are available.

  • Monitor
    Input:  ECG activity may be viewed through the paddles or patient cable. 
    Lead fault:  LEADS OFF message and dashed baseline appear on the monitor if a lead becomes disconnected.  Common mode rejection: greater than 100dB measured as per AAMI standards for cardiac monitors. 
    Heart rate display:  digital read out on monitor from 20 to 280 bpm.

  • Recorder
    Annotations:  time, date, heart rate, event markers, ECG mode, and selected energy. 
    Recorder Mode:  automatically documents events and ECG during defibrillation episodes.  The recorder can be configured to run in either real time or with a 6-second delay. 
    Frequency response:  .5 to 40Hz monitor bandwidth.  .05 to 150 Hz diagnostic bandwidth is selectable.

TOP^