Corometrics 118 Fetal Monitor - Refurbished

(800) 462-8195

Corometrics 118 Fetal Monitor - Refurbished

Web ID: 1721

Condition: Retired

This product is either very difficult to find or no longer available through DRE, please see the list of related items to the right or ask your equipment expert for advice on a similar available product.

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

The Corometrics 118 Fetal Monitor is a twin capable unit that includes one toco and two ultrasound transducers, fetal EKG, event marker and a chart recorder. Data storage and transmission Integrated Maternal Pulse Oximetry NIBP. Extends upon the advanced technology of the Model 116 with integrated maternal pulse oximetry and/or noninvasive blood pressure monitoring.

Features

  • Easy operation
  • Intuitive controls and menus
  • Simplified operation of all functions
  • Controls are tailored for labor and delivery to make operation user friendly
  • Visual/Audible Maternal Alarms
  • The unit provides visual and audible alarms when maternal SpO2, NIBP or pulse rate values exceed user-defined high/low limits
  • Audible alarms can be disabled
  • Individual audio/visual alarms can be temporarily canceled

Specifications

Physical Weight & Dimensions

  • Height: 7 inches (17.8 cm)
  • Width: 16.5 inches (41.9 cm)
  • Depth: 17 inches (43.2 cm)
  • Weight: Approx. 34 lbs (15.5 kg)

Power Requirements

  • Voltage: 100/120/220/240 VAC selectable
  • Frequency: 50/60 Hz.
  • Power consumption: 100 Watts (max)

Direct ECG

  • Technique: peak detecting beat-to-beat cardiotachometer.
  • Heart rate counting range: 30-240 BPM.
  • Heart rate resolution: ±1 BPM.
  • Artifact elimination: switch selectable±25 BPM artifact rejection.
  • Input signal range differential: 15 µV to 2 mV peak-to-peak (fetal), 100 µV to 2 mV peak-to-peak (maternal).
  • Common mode: 20 VDC (max.).
  • Preamplifier bandwidth: 1.0 Hz to 100 Hz.
  • Common mode rejection: >120 dB balance. With patient cable: >110 dB 5k RA or LA unbalance.
  • Input equivalent noise: < 10 µV peak-to-peak.
  • Input impendence: differential mode > 10 MΩ, common mode > 20 MΩ.
  • Main frequency rejection: >40 dB.
  • Leakage current: < 10 µA at 120 VAC.
  • Isolation mains to patient connection: >2kV.
  • Leg plate tester: simulated r-wave at 120 ±1 BPM.

Ultrasound

  • Technique: pulsed doppler with autocorrelation processing.
  • Pulse repetition frequency: 4kHz in single ultrasound mode, 2 kHz in dual ultrasound mode.
  • Pulse duration: 92µs.
  • Center frequency: 1.151 MHz
  • Intensity: <10 mW/cm2
  • Heart rate counting range: 50-210 BPM.
  • Leakage current: <10 µA at 120 VAC.

Pressure

  • Range: 0-100 mmHg
  • Tocotransducer: 0-100 relative units (0-120 grams)
  • Excitation voltage: +4.0 VDC
  • Resolution: 1 mmHg
  • Bandwidth: tocotransducer DC to 0.5 Hz, strain gauge DC to 3 Hz
  • Zero temperature drift: < 0.1 mmHg/°C (electronics only).
  • Leakage current: <10 µA at 120 VAC (transducer isolated).

Recorder

  • Uterine activity: chart width: 4 cm, scaling: 25 mmHg/cm (3.33 kPa/cm), range: 0-100 mmHg, resolution: 1 mmHg.
  • Chart drive: speeds: 1, 2 and 3 cm/min; accuracy: ±1%.
  • Heart rate scale: domestic - chart width 7 cm, scaling 30 BPM/cm, range 30-240 BPM, resolution 1 BPM; international: chart width 8 cm, scaling 20 BPM/cm, range 50-210 BPM, resolution 1 BPM