  • Up to ten vital parameters in one system with central station capability
  • Battery back up
  • Interactive, user-friendly data card technology
  • ESI suppression, pacer rejection and defibrillator protection

Specifications

Physical Weight & Dimensions

  • Height: 16.5 cm (6.5 in)
  • Width: 34.3 cm (13.5 in)
  • Depth: 30.5 cm (12 in)
  • Weight: 5.5 Kg (12 lbs) typical

ECG

  • Number of Leads: 3 or 5 lead wire cables
  • Lead Selection: I, II, III, a VR, aVL, aVF, V (5 wire cable)
  • Gain Selection: ½mV, 1mV, 2mV, 4mV, 8mV full scale (all± 2%)
  • Frequency Response: Diagnostic: 0.05-100 Hz (-3dB); Monitor: 0.5-40 Hz (-3 dB)
  • Sweep Speed: 6.25, 12.5, 25, 50 mm/sec

Heart Rate Display Alarms

  • User Selectable Choices: Primary: ECG; Choices: ECG, Pleth, ART, PA, PW
  • Range: 20-300 BPM
  • Update Time: Every beat
  • Accuracy: ± BPM
  • Alarm Limit Range: 80-300 BMP (high); 20-150 BPM (low)

Non Invasive Blood Pressure

  • Technique: Oscillometric during inflation
  • Max Inflation Pressure: Adult 300 mmHg; Neonate: 150mmHg
  • Measurement Cycle: <40 second average (0-300 mmHg: adult cuff)
  • Automatic Measurement Cycles: 1, 3, 5, 15, 30, 60 minutes

Invasive Blood Pressure (Optional)

  • Number of Channels: 4
  • User Selectable Labels: ART, PA, ICP, CVP, PW
  • Isolation: 4kV RMS, 5.5 kV peak from ground related currents
  • Pressure Range: -10 mmHg to 300 mmHg
  • Zero Offset: ±150 mmHg
  • Probe Impedance: 300-2500 input ohms
  • Frequency Response: 0-20 Hz
  • Input Impendence: 820K @ 20 Hz; >100M @ DC
  • Transducer Sensitivity: 5µV/V/mmHg
  • Zero: Auto push button
  • Drift: ±1 mmHg
  • Accuracy: ±1 mmHg
  • Scale Range: -10 to 10, 0 to 20, 0 to 40, 0 to 60, 0 to 100, 0 to 150, 0 to 200, 0 to 300
  • Display: Systolic, diastolic, mean pressure

Impendance Respiration

  • Rate: 0-150 breaths/minute
  • Resolution: 1 breath/minute
  • Accuracy: > of ±1% or ±1 breath/minute

Alarms

  • Alarm Indication: Audible, visual
  • Alarm Silence: Yes (2 minutes, or permanent)

Temperature (Optional)

  • Number of Channels: 2
  • Range: 20-45° C (68-113° F); ±0.5° C (20-34° C); ±0.2°C (34-41°C); ±0.5°C (41-45°C)
  • Display Resolution: 0.1°C or 0.1°F

Oxygen Saturation

  • Averaging Time: Selectable 3-21 seconds
  • Search Time: Selectable, 1-40 seconds
  • Waveform: User selectable
  • Gain: User selectable
  • Audio: Beep with each detected pulse, pitch varies with oxygen saturation
  • Range: 1-99%
  • Resolution: 1%
  • Accuracy: ±2% (70-99%); ±3% (50-70%)
  • High Saturation Alarm: Range: off, 70-99%
  • Low Saturation Alarm: Range: off, 1-99%

Accessories & Supplies

Web ID Compatible ECG Cables, Leads, & Electrodes Part #  
12590 3-Lead ECG Trunk Cable