DRE is now Avante Health Solutions

Visit our new website by using the "Proceed to new URL" button.

DRE Echo CO2 - Capnography Monitor

(800) 462-8195

DRE Echo CO2 - Capnography Monitor

Web ID: 2150

Condition: Factory New

Part Number: 60129SCRS - with Sidestream Respironics CO2

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

Effective Capnography Monitoring ( EtCO2 ).

The DRE Echo Capnography Monitor provides cost-effective EtCO2 monitoring with accuracy and durability. The Echo CO2 ensures effective monitoring for intubated and non-intubated patients for continuous long-term monitoring. It is tailored for mechanically ventilated and non-intubated patients. Nellcor SpO2 available as an additional option.

Capnography is outpacing Pulse Oximetry as the new standard of care in outpatient surgery procedures. Click to see the dramatic difference between CO2 and SPO2.

Features

  • 5.7 inch high resolution display for easy reading
  • Lightweight, portable design and user-friendly interface for easy operation
  • Flexible configurations to meet different clinical needs
  • Optional — Nellcor SpO2
  • PR measurement
  • Respironics Loflo sidestream or CAPNOSTAT 5 EtCO2 mainstream measurement
  • Powerful storage capacity
  • Real-time parameters measurement display with trend table for easy reviewing
  • Built-in rechargeable Lithium-ion battery for 10 hours continuous working
  • Bi-directional communications with central station by wired or wireless network
  • Nurse call
  • Suitable for adult, pediatric and neonatal patient

Specifications

General

  • Size: 200.8mm (L) × 241mm (H) × 189 mm (D)
  • Weight: 3 kg
  • Display: 5.7 inches
  • Color TFT resolution: 640X480
  • Power Supply: 100-240 VAC, 50/60HZ
  • Pmax=70VA FUSE T 1.6AL
  • Battery Type: Lithium-ion
  • Voltage: 14.8 V DC Capacitance: 4,400 mAh
  • Working period: Color TFT 480min
  • Rechargeable period: < 360min
  • Thermal Recorder: (Optional)
  • Record Width: 48 mm
  • Paper Speed: 25mm/s

SpO2 (Standard)

General

  • Measuring Range: 0 ~ 100 %
  • Alarm Range: 0 ~ 100 %
  • Resolution: 1 %

Accuracy

  • Adult (including Pediatric):
    • ± 2% (70%~100% SpO2)
    • Undefined 0~70% SpO2
  • Neonate:
    • ± 3% (70%~100% SpO2)
    • Undefined (0~70% SpO2)
  • Pulse Rate:
    • Measuring and Alarm Range: 30 ~ 300bpm
    • Resolution: 1bpm
    • Accuracy: ± 3bpm
    • Data update period: 2s
    • ISO 9919

SpO2 (Option, by Nellcor OxiMax)

General

  • Measuring Range: 1 ~ 100 %
  • Alarm Range: 1 ~ 100 %
  • Resolution: 1 %

Accuracy

  • Adult (including pediatric):
    • ±2% (70%~100% SpO2)
    • Undefined: (0~70% SpO2)
  • Neonate:
    • ±3% (70%~100% SpO2)
    • Undefined: (0~70% SpO2)
  • Pulse Rate:
    • Measuring and Alarm Range: 20~300 bpm
    • Resolution: 1 bpm
    • Accuracy: 3 bpm

Respironics CO2

Mainstream

  • CO2 Measurement Range:
    • 0 to 150 mm Hg,0 to 19.7%,
    • 0 to 20 kPa (at 760 mm Hg)
  • Rise Time:
    • Less than 60 ms Adult/Infant
    • Reusable or Single Patient Use
    • Airway Adapter
  • CO2 Resolution:
    • 0.1 mm Hg — 0 to 69 mm Hg
    • 0.25 mm Hg — 70 to 150 mm Hg
  • CO2 Accuracy:
    • 0-40 mm Hg — ±2 mmHg
    • 41 - 70 mm Hg — ± 5% of reading
    • 71 - 100 mm Hg — ± 8% of reading
    • 101 - 150 mm Hg — ±10% of reading
    • Above 80 bpm — ±12% of reading
  • CO2 Stability:
    • Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
    • Long Term Drift: Accuracy specification will be maintained over a 120-hour period
  • Respiration Rate Accuracy: ±1 breath
  • Calibration: No routine user calibration required. An airway adapter zero is required when changing to a different style of airway adapter
  • Water Resistance: IPX4-Splash-proof (sensor head only)
  • Shock Impact:
    • EN60068-2-6 Sinusoidal Vibration
    • EN60068-2-27 Shock
    • EN60068-2-64 Random Vibration
    • Able to withstand repeated 6 foot drops onto tiled floor while operating

Sidestream

  • Sample Rate: 50 m/L per minute
  • CO2 Measurement Range:
    • 0 - 150 mm Hg, 0 to 20 kPa (at 760 mm Hg).
    • Barometric Pressure supplied by Host
  • CO2 Resolution:
    • 0.1 mm Hg 0 to 69 mm Hg
    • 0.25 mm Hg 70 to 150 mm Hg
  • CO2 Accuracy:
    • 0 - 40 mm Hg ±2 mm Hg
    • 41 - 70 mm Hg ±5% of reading
    • 71 - 100 mm Hg ±8% of reading
    • 101 - 150 mm Hg ±10% of reading
    • Above 80 bpm ±12% of reading
  • CO2 Stability:
    • Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
    • Long Term Drift: Accuracy specification will be maintained over a 120-hour period
  • Respiratory Rate Range: 2 to 150 bpm
  • Respiratory Rate Accuracy: ±1 breath
  • Sample Cell/Filter: Proprietary single patient use sample cell and inline filter are integrated with the sample line which eliminates contamination of the internal system.
  • Nasal Sampling Kits for Non-intubated Patients:
    • Adult, pediatric and infant nasal CO2 sampling, nasal CO2 sampling and O2 delivery
    • Adult and pediatric nasal/oral CO2sampling, nasal/oral CO2 sampling and O2 delivery
  • On-Airway Adapter KITS for Intubated Patients:
    • Adult/Pediatric with and without dehumidification tubing
    • Pediatric/Infant, low dead space, with and without dehumidification tubing
    • Taper meets ISO 5356-1
  • Sample Kit Hours of Use:
    • Nasal Cannula (all styles) — up to 12 hours
    • On-Airway Adapter Kits without dehumidification tubing — up to 12 hours
  • Sample Cell Detection: Insertion automatically turns sampling pump on. Removal automatically turns sampling pump off.
  • Water Resistance IPX4: Splash-proof (when sample cell is inserted in sample cell receptacle)
  • Shock ImpactIEC TR 60721-4-7 Class 7M3 (designed to withstand environments subject to significant vibrations or high shock levels)
    • EN60068-2-27 Shock
    • EN60068-2-64 Random vibration

Specifications subject to change without notice.

Accessories & Supplies

Web ID SPO2 Sensors and Probes Part #  
13163 Adult Hard Shell SpO2 Sensor P02796
13236 SpO2 Extension Cable P02796EXT
13237 Y-Site SpO2 Probe P02796Y
Web ID CO2 Cannulas Part #  
513 CO2 Sampling Cannula with Split O2 Delivery P03082D
13238 Sampling Line w/ Dehumidification Tube P01933S
13239 Sidestream LoFlo CO2 Module P02211
Web ID Equipment Stands and Carts Part #  
13235 Rolling Stand 6600RD
13240 Mounting Plate for Rolling Stand 60129MP
13241 Bracket for IV Pole Mount 60129IVB
Web ID Patient Monitor Paper Part #  
13242 Printer Paper 60129PP