DRE is now Avante Medical Surgical

Visit our new website by using the "Proceed to new URL" button.

DRE Signal SP Vital Signs Monitor - Refurbished

(800) 462-8195

DRE Signal SP Vital Signs Monitor - Refurbished

Web ID: 1935

Condition: Retired

This product is either very difficult to find or no longer available through DRE, please see the list of related items to the right or ask your equipment expert for advice on a similar available product.

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

Ultra-lightweight and portable with up to 6 hours of battery life.

Measures Non-Invasive Blood Pressure, SpO2, Stat-Temperature.

SureTemp™ Thermometer provides an accurate reading within seconds

  • Suitable for adult, pediatric and neonatal patients
  • Nellcor SpO2 technology, and motion tolerant SPO2 and NIPB
  • Adjustable audible and visual alarms
  • Removable and rechargeable battery provides up to 6-hour working time
  • Integrated handle, full mounting solutions and integrated printer

Specifications

Physical Weight & Dimensions

  • Height: 240 mm
  • Width: 170 mm
  • Depth: 170 mm
  • Weight: < 3.5g (including a battery)

Safety

Meets the requirements of IEC60601 series, CE marking according to MDD93/42/EEC

Operation Environment

  • AC Power Req: 100-240VAC (± 10%), SO/60 Hz (± 3 Hz), 70VA
  • Temperature: 0-40°C (SureTemp™: 10-40°C)
  • Humidity: 15%-95%, non-condensing

Patient Range

Adult, pediatric and neonatal

NIBP

  • Method: Automatic oscillometric
  • Operation modes: Manual / Automatic / Continual
  • Auto measure time: Adjustable
  • Measurement unit: mmHg / kPa selectable
  • Measurement types: Systolic, Diastolic, Mean, Pulse Rate
  • Measurement range: Adult: 10-270mmHg; Pediatric: 10-200mmHg; Neonate: 10-135mmHg

Over-pressure

  • Protection: YES
  • Resolution: 1mmHg
  • Alarm: Systolic, Diastolic, Mean, Pulse Rate
  • Pulse Rate Range: 40-240 bpm
  • Pulse Rate Resolution: 1bpm
  • Pulse Rate Accuracy: ± 2bpm

Pulse Rate

Range and accuracy: 20-250 bpm ± 3 bpm, 251-300 bpm unspecified

Typical Measurement

Time: < 12s

SpO2

  • Measurement range: 0-100%
  • Resolution: 1%
  • Accuracy:
    • Adult/Pediatric: ±2% (70-100%)
    • Neonatal: ±3% (70-100%)
    • 0-69% unspecified
  • Alarm range: 0-100%
  • Pulse rate:
    • Range: 20-254bpm
    • Resolution: 1bpm
    • Accuracy: ± 3bpm
    • Alarm range: 20-254bpm
    • Storage Humidity: 10%RH to 95%RH

Nellcor OxiMax SpO2

  • Measurement range: 0-100%
  • Compatible sensors:
    • Nellcor sensor type: OxiMax series MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I, MAX-FAST, MAX-R,OxiCliq A, OxiCliq N, OxiCliq P, OxiCliq I, D-YS, D-YSE, D-YSPD, DS-100A, OXI-AIN and OXI-P/I
    • SpO2 accuracy:
      • 70%-100% ±2%; 0-69% unspecified (sensors: MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I and MAX-FAST)
      • 70%-100% ± 2.S%; 0-69% unspecified (sensors: OxiCliq A, OxiCliq N, OxiCliq P and OxiCliq I)
      • 70%-100% ± 3%; 0-69% unspecified (sensors: D-YS, DS-100A, OXI-A/N and OXI-P/I)
      • 0%-100% ±3.5%; 0-69% unspecified

SureTemp™ Thermometer

  • Measurement range: 25 - 44°C (77-111.2°F)
  • Resolution: In MONITOR mode: 0.1°C (0.2°F)
  • Accuracy:
    • In MONITOR mode: 25-32°C (77-89.6 °F) : ±0 .2°C (±0.3 °F) including 32°C (89.6 °F)