DRE is now Avante Medical Surgical

Visit our new website by using the "Proceed to new URL" button.

DRE Waveline Touch Patient Monitor with Touch-Screen

(800) 462-8195

DRE Waveline Touch Patient Monitor with Touch-Screen

Web ID: 1979

Condition: Factory New

Part Number: 66011B2x (x = additional suffix for varying configurations)

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

User-friendly touch-screen patient monitor measures ECG, SpO2, NIBP, heart rate, temperature and respiration. Offers optional dual IBP, and EtCo2 — Now with Masimo SET® Pulse Oximetry!

The fully-functional Waveline Touch features a 12.1" color TFT display with a touch-screen interface. It features multi-lead ECG monitoring, ST segment monitoring and 72 hours trend data analysis.

Features

  • New - Masimo SET® Pulse Oximetry — ask your DRE representative for details
  • New - Optional Plug-In Anesthesia Gas Module expands measurement capabilities — Anesthesia Gas Module
  • New - Bigger time stamp in upper right corner
  • New - Standby button
  • New - Revised SpO2 pitch tone is still audible when SpO2 rate decreases.
  • New - Volume and Sounds are adjustable and now can be turned off completely
  • New - Option to print Numerical Data Only
  • New - Color of the waveforms can be changed
  • EtCO2 Option allows compliance with ASA
  • Available with optional dual invasive blood pressures, EtCO2, a battery and a built-in recorder.
  • 12.1" color TFT display with a touch-screen interface
  • Multi-lead ECG monitoring
  • Built-in battery (optional)
  • NIBP list display
  • OxyCRG dynamic view display
  • Built-in thermal recorder
  • Networkable with a central monitoring system via a wireless or wired network
  • S-T Segment and 72 hours trend data analysis
  • Compact design convenient for mobile monitoring
  • Drug dose calculations
  • Multi-lingual settings

Specifications

Physical Weight & Dimensions

  • Weight: 5kg
  • Height: ~11 inches
  • Depth: ~6-1/5 inches
  • Width: ~12 3/5 inches

Application

  • Neonatal, pediatric and adult patients

Operation Environment

  • Power Source - External AC power or internal battery
  • AC Power - 100-240VAC, 50/60Hz, <150VA
  • Battery - Rechargeable Lead-Acid; Operating time under the normal condition (1 hour); Operating time after the first alarm of low battery (10 min.)

Temperature

  • Working - 5°-40° C
  • Storage - -20°-65° C

Performance Specifications

  • Display - 12.1" color TFT, Touch Screen
  • Resolution - 800 × 600 pixels
  • Trace - 4, 6 or 8 waveforms ECG (I, II, III, aVR, aVL, aVF, V1-V6), PLETH, RESP, IBPx2, ETCO2
  • Indicator - Alarm indicator; Power indicator QRS beep and alarm sound
  • Trend Time - 72 hours
  • Recorder - Built-in, thermal array, 2 channels
  • Record - Width 48mm
  • Recorder - Paper 50mm
  • Record - Speed 25mm/s, 50mm/s

ECG

  • Input - 3-lead ECG cable and standard AAMI line for connection
  • Input - (Optional) 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice - I, II, III, aVR, aVF, aVL, V
  • Gain Choice - x0.5, x1.0, x2.0
  • Frequency Characteristic - 0.05-40 Hz
  • ECG Waveforms - 7 channels
  • Penetration Voltage - 4000VAC 50/60Hz
  • Sweep Speed - 12.5, 25 and 50 mm/s
  • HR Display Range - 30-300bpm
  • Accuracy - ±1% or ±1bpm, whichever is greater
  • Alarm Limit Range Setting - Upper limit 80-400bpm; Lower limit 20-150bpm

RESP

  • Measure Method - RA-LL impedance
  • Range - 0-120 rpm
  • Accuracy - + 3 rpm
  • Alarm Upper-lower Limit Setting - Upper limit 6-120 rpm, Lower limit 3-120 rpm
  • Sweep Speed - 12.5 and 25 mm/s

NIBP (Suntech)

  • Measuring Technology - Automatic oscillating measurement
  • Cuff Inflating - <30s-0-300 mmHg, standard adult cuff
  • Measuring Period - AVE<40s
  • Mode - Manual, Auto, STAT
  • Measuring Interval in AUTO Mode - 2 min-4 hrs
  • Pulse Rate Range - 30-250 (bpm)
  • Measuring Range - Adult/Pediatric Mode, SYS 40-250 (mmHg), DIA 15-200 (mmHg); Neonatal Mode, SYS 40-135 (mmHg), DIA 15-100 (mmHg)
  • Resolution - 1mmHg
  • Accuracy - Pressure, Maximum Mean error ±5mmHg, Max. Standard deviation 8mmHg
  • Overpressure Protection - Adult Mode 280(mmHg), Neonatal Mode 150 (mmHg)
  • Alarm Limit Setting - SYS 50-240 mmHg, DIA 15-180 mmHg

TEMP

  • Range - 25°-50° C
  • Accuracy - + 0.2 °C 25.0°-34.9° C; + 0.1° C 35.0°-39.9° C; + 0.2° C 40.0°-44.9° C; 0.3° C 45.0°-50.0° C
  • Display Resolution - 0.1° C
  • Alarm Upper-lower Limit Setting - Upper limit 0-50° C, Lower limit 0-50° C
  • Channel - 2 channels
  • Alarm Limit Setting - 10-50° C

Masimo SET SpO2

  • SpO2 Accuracy (non-motion):
  • Adult Pediatric: 70~100 %: ±2 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %,0~69 %: unspecified
  • SpO2 Accuracy (motion):
  • Adult Pediatric: 70~100 %: ±3 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
  • SpO2: ±2 %
  • PR: ±3 bpm
  • Modes:
  • Averaging mode: 2,4,8,10,12,14 and 16 s
  • Sensitivity: Normal, APOD and Maximum
  • PR Accuracy (non-motion):
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Adult Pediatric Neonate: 25~240 bpm: ±3 bpm
  • PR Accuracy (motion):
  • Adult Pediatric Neonate: 25~240 bpm: ±5 bpm
  • Measuring Range:
  • SpO2: 1~100 %
  • PR: 25~240 bpm
  • Perfusion: 0.02~20 %

IBP (Option)

  • Measurement Range - -10-300mmHg
  • Channel - 2 channels
  • Pressure Transducer Sensitivity, 5μV/V/mmHg
  • Impedance Range - 300-3000Ω
  • Transducer Sites - ART, PA, CVP, RAP, LAP, ICP
  • Resolution - 1mmHg
  • Accuracy - ±1mmHg or ±2%, whichever is greater
  • Alarm range - -10-300mmHg

EtCO2 (Option)

  • EtCO2 Option allows compliance with ASA Standards for Basic Anesthetic Monitoring (Effective July 1, 2011)
  • Mode of Sampling - Sidestream
  • Principle of Operation -Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
  • CO2 Measurement Range - 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
  • CO2 Calculation Method - BTPS (Body Temperature Pressure Saturated)
  • CO2 Resolution - 0.1mmHg(0-69mmHg); 0.25mmHg(70-150mmHg)
  • CO2 Accuracy - 0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading; 71-100 mmHg ± 8% of reading; 101-150 mmHg ± 10% of reading; Above 80 breath per minute ± 12% of reading
  • Sampling Rate - 100Hz
  • Respiration Rate - 2-150 bpm
  • Respiration Rate Accuracy - ±1 breath
  • Response Time - <3 seconds - includes transport time and rise time
  • Inspired CO2 Measurement Range - 3-50 mmHg

Accessories & Supplies

Web ID EKG / ECG Cables Part #  
13403 3 Lead ECG Cable P05022
13380 3 Lead ECG Cable With Detachable Leads P05020
13404 5 Lead ECG Cable P05023
13381 5 Lead ECG Cable With Detachable Leads P05021
13382 Adult Snap Electrodes Disposable P00842
Web ID NIBP Hoses Part #  
13389 NIBP Hose P02112
Web ID Blood Pressure Cuffs Part #  
6222 DRE Adult Cloth Blood Pressure Cuff
6234 DRE Small Adult Cloth Blood Pressure Cuff
6224 DRE Child Cloth Blood Pressure Cuff
6232 DRE Infant Cloth Blood Pressure Cuff
6223 DRE Adult Thigh Cloth Blood Pressure Cuff
6233 DRE Large Adult Cloth Blood Pressure Cuff 80031LA
Web ID Equipment Stands and Carts Part #  
13235 Rolling Stand 6600RD
Web ID Patient Monitor Supplies Part #  
13385 Wall Mount P02116
13386 Base Plate 66011MP
13387 Printer 6600P
13388 Power Cord P00100
13383 IBP Cable P02118B2
13384 Disposable IBP Transducer P01858B2
13393 Skin Temperature P05054
13394 Oral/Rectal Temperature P05054O
Web ID Patient Monitor Paper Part #  
3467 Printer Paper for Waveline Pro, EZ, Touch and Plus (5 Rolls) P01620.5
Web ID CO2 Cannulas Part #  
13239 Sidestream LoFlo CO2 Module P02211
13391 Sidestream CO2 Module Bracket P02211B
13405 Mainstream CO2 Modules P02212
13238 Sampling Line w/ Dehumidification Tube P01933S
13392 Tee Adapter for CO2 Sampling (10/pk) P02115
13406 Mainstream Adult/Pediatric Airway Adapter P00300
13407 Mainstream Neonatal Airway Adapter P00301
Web ID SPO2 Cables Part #  
13374 Analog SpO2 Finger Probe P0375AFSEXT
13375 Analog SpO2 Extension Cable P0395EXT
13376 Analog SpO2 Y-Site Probe P02270AY