Operating Temperature: 41°F to 104°F (5°C to 40°C)
Storage Temperature: -40°F to 158°F (-40°C to +70°C)
Operating Humidity / Storage: 5% to 95%, non-condensing
Operating Altitude: 500 mbar to 1060 mbar pressure -1000 ft to 18,000 ft (-304 m to 5,486 m)
Modes
Averaging mode: 2, 4, 8,10,12,14 or 16 seconds
Sensitivity: APOD, Normal and Max
Alarms
Audible and visual alarms for high and low saturation (1% to 100%), pulse rate (25 - 240 bpm), sensor condition, system failure and low battery.
Alarm volume 75db min
Display/Indicators
Data display: %SpO2, pulse rate, alarm status, alarm silenced status, AC power, Signal IQ / pleth bar, perfusion index bar, battery status, no sensor, sensor off
APOD, Norm, Max, FastSat, Trauma: LED
Type: LED
Output Interface
Serial RS-232
Nurse Call
Compliance
EMC Compliance: EN60601-1-2, Class B
Equipment Classification: IEC 60601-1/ UL 60601-1
Type of Protection: Class 1 (on AC power); Internally powered (on battery power)
Degree of Protection - Patient Cable: Type BF-Applied part