Description
Hospital-grade patient monitor measures NIBP, ECG, Resp., Dual Temp, SpO2 & EtCO2 (optional)
The DRE ASM 6000 is a true hospital grade patient monitor that sacrifices nothing in terms of quality. It features a flexible, upgradable and comprehensive software package that provides everything needed in today's demanding healthcare environment. The DRE ASM 6000 has the industry's most clinically proven technology from manufacturers such as Nellcor®, Suntech® and Oridion®.
Measures: Non-invasive Blood Pressure (Suntech®), ECG (3/5 Lead), Respiration, Dual Temperature and SpO2 (Nellcor OxiMax®), and optional EtCO2 (Oridion®)
The DRE ASM 6000 also includes a wireless remote control which operates as far as 20 feet from the monitor. This allows you to control the DRE ASM 6000 from anywhere in your operating room!
Features
- Simultaneous multi-lead ECG monitoring
- Increasing/decreasing pulse tone
- Adult, pediatric and neonate modes
- User-selectable screen formats and colors
- Four-hour battery operation
- Optional ST and Arrhythmia Analysis
- 72 hour / 100 patients advanced trending
- Paperless USB data management
- Built-in dual channel recorder
Specifications
Display
- Type: High resolution color TFT display
- Size: 8.4" (diagonally)
- Resolution: 800 × 600 pixels, excellent viewing for distance and angle
- Parameters: ECG, Respiration, SpO2, Dual NIBP, Temperature & CO2
- Trace Speed:
- 2.5, 25 & 50 mm/sec. for ECG & SpO2
- 6.5, 12 & 25 mm/sec. for Respiration & CO2
Optical Encoder / Keys
- Optical Encoder: Rapid access to all the functions and setting of parameters through single knob
- Hot Keys: Quick action hot keys for alarm acknowledge, NIBP start/stop, recorder start/stop, freeze, stand by, go-to and return to the main screen
ECG Monitoring
- Lead: 3/5 lead with optional ST & Arrhythmia
- Protection: In-built cautery & defibrillator protection
- Pacer Detection: Possible
- HR Range: 20 to 350 bpm (±2 bpm or 2%, whichever is greater)
Respiration (from ECG)
- Principle: Impedance pneumography
Pulse Oximetry (Nellcor Module)
- Tone Variation: Tone variation with change in SpO2 values
- SpO2 Range: 1 - 100%
- Accuracy - Adult:
- 100 - 70% (±2 digits)
- 0 - 69% (unspecified)
- Neonates: 100 - 70% (±3 digits)
- Perfusion Indicator: Bar graph showing signal strength
- Modes: Normal response & fast response
- Alarms:
- Low 0 - 95%
- High 5 - 100%
- PR Range: 20 to 230 bpm
Temperature
- Channel: Dual channel
- Range: 0° C to 50° C (±0.2° C or ±0.4° F)
Probe compatible with YSI 400 series probes (optional)
Non-Invasive Blood Pressure (Suntech Module)
- Principle: Oscillometic
- Display: Systolic, Diastolic & Mean pressure
- Modes: Manual, STAT (continuous 5 min. operation) and automatic (time interval 2 - 90 min. selectable)
- Alarm Limits: Selection possible of r Systolic & Diastolic
- Range: 20 - 250 mmHg
- Accuracy: ±3 mmHg
Capnography (Oridion Module)
Microstream Capnography with display of CO2 waveform and digital values of EtCO2, FiCO2 & RR. CO2 gets priority for calculating Respiration rate.
- EtCO2 Range: 0 - 99 mmHg
- FiCO2 Range: 0 - 20 mmHg
- Gain: 20, 40, 60, 80 mmHg
- Flow Rate: 50 ml/min. (-7.5 / +15 ml/min.)
- Warm-up time: 40 sec. (typically), Units: mmHg / Kpa / Vol%
- RR Range: 0 - 150 bpm
Alarms
Alarm setting for all parameters.
- Patient Alarm: RED flashing with audio and message indicator
- Equipment Alarm: YELLOW flashing with audio and message indicator
- Alarm Suspend: Continuous YELLOW
- Auto Setting: HR / PR, SpO2, RR, Temperature & CO2
- Manual Setting: HR / PR, SpO2, Systolic, Diastolic, RR, Temperature & CO2
- Re-Alarm Facility 3 min. for acknowledge alarm
Trends
72 hrs / 100 patients graphical and tabular trend for HR / PR, SpO2 , RR, Temperature & CO2 & ST.
- NIBP Trend: Last 240 readings (tabular)
- Alarm Recall: Last 12 critical alarms with date & time
Recording (optional)
- Type: In-built dual channel thermal array recorder for recording ECG, plethysmograph, Respiration & CO2 waveforms (any two-user selectable) with numerical data, NIBP & Tabular trend recording
- Modes: Direct / delayed / continuous and record on alarm features
Slave Display
- Dimension: Possible with external monitor
- HxWxD: 235 × 260 × 158 mm
- Weight: 4.5 kg. (with in-built battery & thermal array recorder
Power Supply
- Universal Poser: 95 - 264V AC, 50 160 Hz (±5%)
- Current Rating: 0.7 A (max)
- Battery Type: 14.8V, (4Ah / 8Ah) Lithium ion (in-built)
- Battery Operation: Rechargeable battery for 2/4 hrs of continuous operation under full charge condition (typically)
- Charging Time: 8 hrs. (typically)