DRE Trax Monitor De Signos Vitales

Web ID: 2379

Condición: Factory New

Part Number: 66311x (x = additional suffix for varying configurations

0 Los artículos están en tu carrito. Mirar Carrito ya.

¿Necesita asesoramiento sobre los equipos? ¡Pregunte a Nuestros Expertos!

(502) 244-6333

Descripción

El monitor DRE Trax es muy vers

Especificaciones

General

  • Display: 5.0 inch (Diagonal) color TFT
  • Resolution: 800 × 3(RGB) × 480
  • Trace: 2 waveforms
  • Waveforms: PLETH, ETCO2
  • Trend time: From 1 to 72 hours
  • Indicator:
    • Alarm Indicator
    • Power indicator
    • Pulse beep and alarm sound

NIBP

  • Measuring Technology: Automatic oscillating measurement
  • Cuff Inflating: <30s (0 ~ 300 mmH, standard adult cuff)
  • Measuring Period: AVE<40s
  • Mode: Manual, Auto, STAT
  • Measuring Interval in AUTO Mode: 2 min ~ 4 hrs
  • Pulse Rate Range: 30 bpm ~ 250 bpm
  • Measuring Range: Adult/Pediatric Mode
  • SYS: 40 ~ 250 (mmHg)
  • DIA: 15 ~ 200 (mmHg)
  • Neonatal Mode
  • SYS: 40 ~ 135 (mmHg)
  • DIA: 15 ~ 100 (mmHg)
  • Resolution: 1mmHg
  • Pressure Accuracy: Maximum Mean error: ±5mmHg
  • Maximum Standard deviation: 8mmHg
  • Overpressure Protection: Adult Mode: 280(mmHg)
  • Neonatal Mode: 150 (mmHg)
  • Alarm Limit: SYS: 50 ~ 240 mmHg
  • DIA: 15 ~ 180 mmHg
  • Standards: Meets performance standards of ANSI/AAMI SP10:2002

SPO2

  • ASpO2: Anti-motion SpO2
  • SpO2% Range: 0 ~ 100%
  • SpO2 Accuracy: ±2% (70 ~ 100%,non-motion)
  • ±3% (70 ~ 100%, motion)
  • Pulse Rate Range: 30-250 bpm
  • Pulse Rate Accuracy: ±2 bpm(non-motion), ±3 bpm (motion)
  • Alarm Upper-lower Limit: Upper limit 70 ~ 100%, Lower limit 70 ~ 100%
  • SpO2 Probe: Red light LED wavelength: 660nm±5nm
  • Infrared light LED wavelength: 940nm±10nm
  • Standards: Meets performance standards of EN ISO 9919:2005

Rapid Temperature (Option)

  • Temperature Measurement Range: 30°C to 43°C (86°F to 109°F)
  • Typical Measurement Times
  • Oral (Quick Mode):3-5 seconds (non-fever temps), 8-10 seconds (fever temps):
  • (after insertion into measurement site):
    • Oral (Standard Mode): 6-10 seconds
    • Axillary Mode: 8-12 seconds
    • Rectal Mode: 10-14 seconds
    • Direct Mode (All Sites): 60-120 seconds
  • Approx. 6000 temperature readings
  • Standards: Meets performance standards of EN 12470-3:2000, ASTM E1112:2006
  • Pulse Timer: 60 Second count with a "beep" at 15 seconds, 2 "beeps" at 30 seconds, 1 "beep" at 45 seconds, and 2 "beeps" at 60 seconds
  • Patient Accuracy: A Standard Prediction Mode reading and a Direct Mode reading will differ by less than ±0.2°C (±0.4°F) on 98% of tested patients
  • Batteries:
    • Four "AA" Required.
    • Standard IEC package size.
    • Alkaline --1.5 Volt

EtCO2 (Option)

  • Mode of Sampling: Sidestream or Mainstream
  • Principle of Operation: Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts.
  • CO2 measurement Range: 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
  • CO2 Calculation Method: BTPS (Body Temperature Pressure Saturated)
  • CO2 Resolution: 0.1mmHg (0-69mmHg), 0.25mmHg (70-150mmHg)
  • CO2 Accuracy:
    • 0 ~ 40 mmHg ± 2 mmHg
    • 41 ~ 70 mmHg ± 5% of reading
    • 71 ~ 100 mmHg ± 8% of reading
    • 101 ~ 150 mmHg ± 10% of reading
    • Above 80 breath per minute ± 12% of reading
  • Sampling rate: 100Hz
  • Respiration Rate: 2 ~ 150 bpm
  • Respiration Rate accuracy: ±1 breath
  • Response Time: <3 seconds -
  • includes transport time and rise time
  • Inspired CO2 measurement Range: 3 ~ 50 mmHg
  • Standards: Meets performance standards of ISO/FDIS 21647:2004 (E), ASTM F1456-01, IEC/CDV 60601-2-55

Networking

  • Wired Networking:
    • Industry standard 802.11b/g wired network
    • Frequency Range: 2.412 ~ 2.484 GHz
    • Connected bedside number: Up to 16 bedside monitors
  • Wireless Networking:
    • Up to 100m indoors
    • Industry standard 802.11b/g wireless
    • Supports TCP/IP and UDP/IP Protocols

Power

  • Source: External AC power or internal battery
  • AC Power: 100 ~ 240VAC, 50/60Hz, 150VA
  • Battery: Built-in and lithium Ion rechargeable, 12.6V/5Ah
  • Charge Time: 8 hours
  • Operating Time: 3 hours

Environmental

  • Temperature: Operating: 5 ~ 40 °C
  • Storage: -10 ~ 45 °C
  • Humidity Range: Operating: ≤80 %
  • Storage: ≤80 %

FUSE

  • 3.15A/250V

LCD

  • Display Type: TFT color LCD
  • Size (diagonal): 5.0 inch
  • Active Area: 152.4 (W) × 91.44 (H) mm
  • Color arrangement: RGB-stripe
  • Dot pitch: 0.0635(W) × 0.1905(H) mm
  • Display Mode: Normally white, Transmissive
  • Interface: Digital (TTL)
  • Surface Treatment: Anti-Glare

Touchscreen

  • Type: Four-Wire Analog Resistive Touch Panel
  • Input Mode: Stylus Pen or Finger
  • Connector: FPC
  • Insulation resistance: 25Mω
  • Voltage: 7VDC
  • Chattering: 10ms
  • Transparency: 80%
  • Surface hardness: 3H
  • Durability-surface scratching: Write 100,000
  • Active force: 80gf
  • Knock Test: 1,000,000 times

Suministros y Accesorioss

Web ID Cánula nasal Part #  
513 Cánula de Muestra de CO2 con Entrega de O2 Dividida P03082D

Shop By Category