DRE is now Avante Medical Surgical

Visit our new website by using the "Proceed to new URL" button.

Respironics Lifecare PLV-102 Ventilator - Refurbished

(800) 462-8195

Respironics Lifecare PLV-102 Ventilator - Refurbished

Web ID: 1676

Condition: Retired

This product is either very difficult to find or no longer available through DRE, please see the list of related items to the right or ask your equipment expert for advice on a similar available product.

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

This lightweight ventilator is well suited for transport and portable applications. Controlled by a microprocessor, the piston-driven PLV-102 delivers volume ventilation in a range of settings. With features that include three power source options, it is ideal for use in the sub-acute care, nursing home, or home settings.

Specifications

Physical Weight & Dimensions

  • Height: 9 inches (22.9 cm)
  • Width: 12.25 inches (31.1 cm)
  • Depth: 12.25 (31.1 cm)
  • Weight: 28.9 lbs (13.1 kg)

Power

  • 120 VAC 50/60 Hz
  • 12 VDC External Battery
  • 12 VDC Internal Battery

Front Panel Controls

  • Mode Selection: Control, Control + Sigh, Assist/Control, Assist/Control + Sigh, SIMV
  • Tidal Volume: 0.05 to 0.20 ± 0.02 liters; 0.20 to 3.00 liters ± 10%
  • Sign Volume: 150% of Set Tidal Volume
  • Sign Frequency: 1 Sign Breath every 100 Tidal
  • Breath Rate: 2 to 35 BPM ± 0.5; 36 to 40 BPM ± 2
  • Inspiratory Flow Rate (Peak flow rate during inspiratory time): 10 to 120 LPM
  • Battery Test Switch: Test Internal/External Battery
  • Main Power Switch: On/Off, Recharge
  • DC Circuit Breaker: Push/Pull Actuation
  • AC Circuit Breaker: Push/Pull Actuation
  • Airway Pressure Limit: 5 to 100 cm H2O ± 5
  • Sensitivity: +18 cm H2O ± 1 to at least -6 cm H2O
  • Low Pressure Alarm: 2 to 40 cm H2O ± 2; 41 to 50 cm H2O ± 5
  • Manual Sigh: Push to Activate One Breath
  • Oxygen Percent (See Operators Manual for Oxygen Delivery Limitations): 21% to 90% ± 6%
  • Oxygen Sensor - SIMV Operation (See Operators Manual for Oxygen Delivery Limitations): 21% to 90% ± 10%
  • 30 Second Alarm Silence: Push to Activate

Front Panel Visual Indicators

  • Manual Sigh: Yellow Indicator Light
  • Tidal Volume: Digital Display
  • Rate: Patient Breaths Per Minute - Digital Display
  • I:E Ratio: Digital Display
  • Inspiratory Flow Rate: Digital Display
  • Increase Inspiratory Flow: Red Indicator Light
  • 120 VAC Power Source: Green Indicator Light
  • Patient Airway Pressure: -10 to 100 cm H2O ± 5 (Pressure waveform is sinusoidal)
  • Internal DC Power Source: Amber Indicator Light
  • External DC Power Source: Yellow Indicator Light
  • Assisted/Spontaneous: Green Indicator Light
  • 15-Second Delay (Low Pressure Alarm): Green Indicator Light
  • Oxygen Percent: 50% and Greater Marked Red
  • 30-Second Alarm Silence: Yellow Indicator Light/Push Button
  • Alarm Code Display (in Flow Window): Digital Display

Alarms

  • Low Pressure: Visual/Continuous Audible
  • High Pressure: Intermittent Audible
  • Apnea: Visual/Continuous Audible
  • Inverse I:E Ratio: Visual Warning
  • Increase Inspiratory Flow: Visual Warning
  • Reverse Battery Connection: Continuous Audible
  • Low Internal Battery: Visual/Continuous Audible
  • Low External Battery: Visual/Continuous Audible
  • Power Failure: Continuous Audible
  • Unit Malfunction: Intermittent Audible
  • Power Source Change: Continuous Audible
  • Oxygen: Continuous Audible