DRE is now Avante Medical Surgical

Visit our new website by using the "Proceed to new URL" button.

Sarns 7000 Heart Lung Bypass - Refurbished

(800) 462-8195

Sarns 7000 Heart Lung Bypass - Refurbished

Web ID: 271

Condition: Refurbished

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

The Air-Shields TI-100 Transport Incubator is available refurbished from DRE.

Features

  • Intended for transport of high risk, premature, low birth-weight, or critically ill newborns.
  • It provides a means to control air temperature, oxygen concentrations, and relative humidity.
  • A double wall hood permits full visibility and provides an effective thermal barrier from the environment.

Specifications

Dimensions (with stand)

  • Height: 46.75 inches (118.7 cm) maximum, 34.75 inches (88.3 cm) minimum
  • Width: 22.25 inches (56.5 cm)
  • Length: 40.25 inches (102 cm)

Power Requirements

  • AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
  • DC: 12 VDC, 115 Watts

Internal Battery

  • Type: gel type, sealed and rechargeable.
  • Voltage: 12 VDC nominal
  • Quantity: 2
  • Charging Time: two batteries take 20 minutes to recharge from fully discharged.
  • Life expectancy: minimum 200 full charge/discharge cycles.

Chassis Leakage Current

100µA or less

Alarms

  • High Temp: Actuates at displayed incubator temperature of 39° ± 0.5 °C.
  • Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
  • Power Failure: Actuates if main AC power fails and no DC back-up power is present.
  • Heater Temp: Actuates if heater temperature exceeds 77°C nominal.
  • Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.

Temperature

  • Set Point Range: 21.5°C ± 1.0°C to 38 °C
  • Rise Time: 30 minutes
  • Variability: 0.2°C
  • Overshoot: 1.5°C maximum
  • Uniformity: 1°C
  • Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1°C
  • Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5°C
  • Observation Light Output: 10-foot candles at mattress level
-------------------------------------------------

The Sarns 7000 Heart Lung Bypass is available refurbished from DRE.