DRE is now Avante Health Solutions

Visit our new website by using the "Proceed to new URL" button.

Tidalwave 715Sp Handheld Monitor with SP O2 - Refurbished

(800) 462-8195

Tidalwave 715Sp Handheld Monitor with SP O2 - Refurbished

Web ID: 405

Condition: Retired

This product is either very difficult to find or no longer available through DRE, please see the list of related items to the right or ask your equipment expert for advice on a similar available product.

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

DRE offers the Respironics Novametrix Tidal Wave model 715Sp portable handheld pulse oximeter and CO2 monitor with SpO2. The Tidal Waves are the only handheld, mainstream capnographs available today, and provide immediate feedback on the patient's airway condition and ventilatory status. Based on the same technology as Respironics Novametrix bedside capnographs.

Three Optional Configurations

  • Tidal Wave Model 610 Handheld Mainstream Capnograph
  • Tidal Wave Model 615 Handheld Capnograph with Mainstream and Sidestream Sampling
  • Tidal Wave Sp Model 710 Handheld Mainstream Capnograph/Pulse Oximeter Tidal Wave Sp
  • Tidal Wave Sp Model 715 (described on this page) Handheld Mainstream Capnograph/Pulse Oximeter with Sidestream Sampling Capabilities

Features

  • Built in Pulse Oximeter
  • Hand-held Mainstream and Sidestream Capnograph
  • 30-minute on-screen graphic trend available
  • Visual and audible "no respiration" alert
  • User replaceable disposable and rechargeable battery packs eliminate the need to wait while batteries are charging (standard AA lithium: 5.5 hours of life)
  • No calibration required for CO2

Specifications

Capnograph

Principle of Operation

Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.

Initialization Time

Capnogram within 15 seconds, full specifications within 60 seconds.

Calibration

No calibration required.

Capnostat® Mainstream CO2 Sensor

  • Size: 1.30 × 1.67 × 0.85 inches (3.30 × 4.24 × 2.16 cm)
  • Weight: 0.64 oz (18 grams), cable excluded
  • Shock Resistance: Withstands repeated six foot drops to a hard surface

CO2 (Carbon Dioxide)

  • Range: 0 - 150 mmHg
  • Accuracy:
    • ±2 mmHg (for 0 - 40 mmHg),
    • ±5% of reading (for 41 - 70 mmHg),
    • ±8% of reading (for 71 - 150 mmHg)
  • Resolution: 1 mmHg
  • Response Time: 60 ms
  • Units: mmHg, kPa, or %
  • Compensations: N2O & O2 (selectable), Pb (automatic)
  • Capnogram: Selectable sweep speed

Respiratory Rate

  • Accuracy: ±1 br/min
  • Range: 0 - 150 br/min
  • Resolution: 1 br/min

Sampling Flow Rate (Model 715 only)

  • 180 ml/min (for non-intubated patients)
  • Response time is less than 200 ms at 180 ml/min flow

Pulse Oximeter

Principle of Operation

Red/Infrared absorption.

SpO2 (Oxygen Saturation)

  • Range: 0-100%
  • Accuracy: ±2% SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
  • Resolution: 1%
  • Averaging: Menu selectable 2 or 8 seconds

Pulse Rate

  • Range: 30 - 250 bpm
  • Accuracy: ±1% of full scale
  • Resolution: 1 bpm
  • Averaging: 8 seconds

Sensors

Reusable Y-Sensor (can be sterilized and used with all patient populations); reusable adult finger sensor

Plethysmogram

Pulsatile waveform

General

Alerts

  • Limits:Adjustable limits for ETCO2, SpO2, Respiratory Rate and Pulse Rate. No Respiration Detected adjustable timeout.
  • Audio: Adjustable volume, two minute silence or off (LED indicators)
  • Visual: On-screen and LED

Display

  • Type: Dot matrix, backlit (LCD)
  • Size: 2.2 × 1.1 inches

Trending Capabilities

  • Memory: 24 hr., battery backed
  • Format: 30 minute on-screen or printed in trend or tabular mode. Download data to personal computer for use with NovaCARD software

Communication Output

RS232 to Seiko DPU-414 printer, NovaCARD™ software or computer

Physical

  • Size (height, width, depth): 8 × 3.75 × 1.875 inches (20 × 9.5 × 4.8 cm)
  • Weight: 793 grams (28 oz.)

Electrical

  • External Supply (optional)
  • Input: 120 VAC 60 Hz, or 220 VAC 50 Hz
  • Battery: Nickel-Metal-Hydride (NiMH): 4.5 hour life
  • Battery: Standard AA lithium: 5.5 hour life

Environmental

  • Operating temperature: 0 to +40°C (32 - 104°F)
  • Operating humidity: 0 to 90% RH non-condensing.