DRE is now Avante Medical Surgical

Visit our new website by using the "Proceed to new URL" button.

Tuttnauer 2540 EL Fully Automatic Benchtop Sterilizer

(800) 462-8195

Tuttnauer 2540 EL Fully Automatic Benchtop Sterilizer

Web ID: 1613

Condition: Factory New

0 items are in your cart. View cart now.

Need Advice? Ask Our Experts!

(800) 462-8195

Description

With higher capacity and better performance, Tuttnauer's EL Laboratory line of fully automatic, microprocessor controlled, bench-top, steam sterilizers provides safe, economical and effective sterilization. Integrated reservoir for water feed, with water level control and steam recirculation saves on demineralized water consumptionand enables the autoclave to be independent of a water supply. External heating elements make the sterilizereasy to clean.

Sterilization Applications

  • Liquids such as nutrient media and buffer solutions
  • Solid items such as pipettes, tubes and filters
  • Glassware and plastic articles
  • Waste sterilization

The various applications can be carried out using the autoclave's standard programs. Additional options ensure more control over results and optimal, safe and repeatable sterilization processes. For special applications, our experienced team will customize a suitable method for you.

Features

  • The sterilizer microcomputer control system provides high reliability and safe operation.
  • 6 individual preset programs.
  • Pre-programmed parameters remain programmed until altered.
  • Sterilization Initiation - the PT-100 temperature sensor in a reference vessel ensures that the set sterilization temperature is actually attained when sterilization starts.
  • Demineralized water level for steam generation is automatically regulated.
  • Exhaust steam is cooled down before leaving the autoclave.
  • LCD Displays pressure, temperature, time and alerts.
  • Heat-resistant plastic covering on lid prevents contact with hot parts and protects from burning.
  • Temperature and pressure regulation
  • Pre-prepared for cooling - The autoclaves are prepared for the retrospective fitting of cooling equipment; both electronics and cooling coils can be fitted without difficulty.
  • Validation ports are available for external temperature sensors for validation and qualification purposes.
  • RS 232 interface

Safety Features

  • Pressure-activated door lock prevents the door from opening under pressure.
  • Temperature-dependent door lock - the system prevents the door from opening until the temperature drops below a specified level, thereby preventing delayed boiling
  • Door lock safety switch prevents the sterilization program from being started if the door has not been closed properly
  • Water level safety device prevents the autoclave from being heated up in the case of insufficient water
  • Excess temperature safety device switches off the heating elements if too high a temperature is attained.
  • Over-pressure valve releases over pressure in order to prevent high pressure risk.

Specifications

General

  • Chamber Dimensions Ø × depth in mm: 250 × 420
  • Chamber Volume in Liters: 23
  • External Dimensions: 365 (H) × 510 (W) × 545 (D) mm
  • Heating Performance: 220KW
  • Voltage in Volt 50/60 Hz: 220-240
  • Internal bottom-space (W x D) in mm: 170 × 420