Description

Features

  • Monopolar and Bipolar Outputs
  • Audible and Visual Arms, including REM™ Fault Indicator
  • REM™ (Return Electrode Monitoring) (Standard)
  • 300 Watts Output - Cut, 120 Watts - Coagulation, 70 watts - Biopolar
  • Monopolar Foot Switch Included
  • Bipolar Foot Switch (Optional)
  • Rolling Carts (Optional)

Specifications

Physical Weight & Dimensions

  • Dimensions: 8 × 11 × 17 inches (15 × 28 × 3 cm)
  • Weight: 15 lbs (6.4 kg)

Output Characteristics

  • Cut: 2400 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 200 Nominal Power at rated load watts, 1.8 @ 100 W Crest Factor at rated load typical
  • Blend 1: 2800 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 175 Nominal Power at rated load watts, 2.7 @ 100 W Crest Factor at rated load typical
  • Blend 2: 3200 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 150 Nominal Power at rated load watts, 3.3 @ 100 W Crest Factor at rated load typical
  • Blend 3: 3600 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 125 Nominal Power at rated load watts, 4.6 @ 100 W Crest Factor at rated load typical
  • Coag: 5000 Max (open circuit) P-P Voltage, 300 Rated Load ohms, 75 Nominal Power at rated load watts, 8.0 @ 50 W Crest Factor at rated load typical
  • Bipolar: 800 Max (open circuit) P-P Voltage, 100 Rated Load ohms, 50 Nominal Power at rated load watts, 1.8 @ 40 W Crest Factor at rated load typical

Output Waveform

  • Cut: 500 kHz Sinusoid
  • Blend 1: 500 kHz bursts, 50% Duty Cycle, 31 kHz Recurrence
  • Blend 2: 500 kHz bursts, 37.5% Duty Cycle, 31 kHz Recurrence
  • Blend 3: 500 kHz bursts, 25% Duty Cycle, 31 kHz Recurrence
  • Coag: 500 kHz bursts, 31 kHz Recurrence
  • Bipolar: 500 kHz bursts, Unmodulated Recurrence

Low Frequency Leakage (50/60 Hz)

  • Normal polarity: intact chassis ground: < 10 mA, ground open; < 100 mA.
  • Reverse polarity: ground open: < 100 mA. Sink current, 140V applied, all inputs; < 150 mA. Chassis source current, ground open; < 110 mA

High Frequency Risk Parameters

Bipolar and Monopolar RF leakage current: < 150 milliamperes

REM™ Contact Quality Monitor

  • Measurement frequency: 140 kHz ± 10 kHz.
  • Measurement current: 1.5 mA ± 0.5 mA

Input Power Source

Nominal input line voltage is 120 VAC with a full range of 85-135 VAC. Frequency line varies from 45 to 65 Hz.

  • Idle: 0.5 A (Max. Current), 60 W (Max. Power)
  • Cut: 5.0 A (Max. Current), 600 W (Max. Power)
  • Coag: 30 A (Max. Current), 300 W (Max. Power)
  • Bipolar: 1.5 A (Max. Current), 180 W (Max. Power)