Description

Features

The Force 1C generator features a compact design suitable for the requirements for outpatient and office-based surgery, and the performance requirements for a full range of in-patient general surgery applications.

Digital Electrosurgical Unit suitable for all uses:

  • Outpatient procedures to full range surgical applications.
  • Monopolar and Bipolar outputs.
  • Audible and visual alarms.

Specifications

Physical Weight & Dimensions

  • Dimensions: 6 × 11 × 17 inches (15 × 28 × 43 cm)
  • 5 lbs (6.8 kg)

Output Waveform

  • Cut: 516 kHz sinusoid;
  • Blend 1: 516 kHz of sinusoid at 70% duty cycle recurring at 31 kHz;
  • Blend 2: 516 kHz bursts of sinusoid at 45% duty cycle recurring at 31 kHz;
  • Blend 3: 516 kHz bursts of sinusoid at 20% duty cycle recurring at 31 kHz;
  • Coag: 516 kHz damped sinusoidal bursts with a repetition frequency of 31 kHz;
  • Micropolar: 516 kHz sinusoid, unmodulated.

Output Characteristic

  • Cut: 2400 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 200 Nominal Power at rated load watts; 1.8 @ 100 W Crest Factor at rated load typical
  • Blend 1: 2800 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 175 Nominal Power at rated load watts; 2.7 @ 100 W Crest Factor at rated load typical
  • Blend 2: 3200 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 150 Nominal Power at rated load watts; 3.3 @ 100 W Crest Factor at rated load typical
  • Blend 3: 3600 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 125 Nominal Power at rated load watts; 4.6 @ 100 W Crest Factor at rated load typical
  • Coag: 5200 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 75 Nominal Power at rated load watts; 8.0 @ 50 W Crest Factor at rated load typical
  • Micropolar: 900 Max. (open circuit) P-P voltage; 100 Rated Load ohms; 50 Nominal Power at rated load watts; 1.8 @ 40 W Crest Factor at rated load typical

Input Power Source

(Max. operating range: 85-135 or 170-270 VAC; Nominal operating range: 110-120 or 220-240 VAC)

  • Idle: 0.5 A Max. Current; 60 W Max. Power
  • Cut: 5.0 A Max. Current; 600 W Max. Power
  • Coag: 3.0 A Max. Current; 400 W Max. Power
  • Micropolar: 2.0 A Max. Current; 180 W Max. Power

Line Regulation

Between 85-135 volts or 170-270 volts input, output power into nominal load will vary no more than 20% or 5 watt, whichever is greater.

REM Contact Quality Mon.

Measurement frequency: 140 kHz ± 20 kHz. Measurement current: 3.0mA, maximum.

Low Frequency Leakage

Source current nominal polarity, intact chassis ground: <10 µA, ground open: <50 µA. Source current reverse polarity, ground open: <50 µA. sink current, 135 VAC applied, all inputs: <20 µ A. Chassis source current, ground open: <100 µA

Low Frequency Leakage

Micropolar RF leakage current: <150 mA. Monopolar RF leakage current: <150 mA